English–DPI610A Instruction Manual | 11

5. Press the RUN key (F1) and operate the hand-pump, continue until the switch operates:

6. Carefully open the release valve and allow airspeed to decrease until the switch operates 

again. The display shows the operating airspeeds and hysteresis.

7. Record the displayed operating data.
8. Allow the airspeed to decrease until it is below 80

 

kts.

9. Slowly open the manual let-down valve to return system to zero (ground level).
10. To perform another switch test, press RUN (F1) and repeat the above procedure.
11. To cancel switch test mode, press TASK.

2.3.2  Method (Non-accessible Switch Contacts)

When the airspeed switch contacts cannot be connected to the instrument, an external indicator 

or annunciator must be used as indication of airspeed switch operation. The user, on receipt of 

an indication of switch operation, presses the MANUAL OPERATION (F2) key:

1. Connect the instrument to the pitot system as shown in Figure

 

4. Ensure that the static 

system is vented to atmosphere before starting the test.

2. Ensure that positive pressure is selected and close the release valve.
3. Close the manual let-down valve.
4. Zero the airspeed reading by pressing the ZERO key.
5. Press the RUN key (F1) and operate the hand-pump, continue until the switch operates and 

immediately press (F2).

6. Carefully open the release valve and allow the airspeed to decrease until the switch operates 

again and immediately press (F2). The display shows the operating airspeeds and 

hysteresis.

7. Record the displayed operating data.
8. Allow the airspeed to decrease until it is below 80

 

kts.

9. Slowly open the manual let-down valve to return system to zero (ground level).
10. To perform another switch test, press RUN (F1).
11. To cancel switch test mode, press TASK.

2.4  Airspeed Leak Tests

Select TASK, AERO:

kts

kts

kts

SWITCH TEST

CONTACT STATE

AIRSPEED

RUN

MANUAL 

OPERATION

STOP

MANUAL 

OPERATION

STOP

AIRSPEED

AIRSPEED

CONTACT STATE
CLOSED AT 

550 

kts

OPENED AT 

525 

kts

HYSTERESIS 25 

kts

SWITCH TEST

SWITCH TEST

CONTACT STATE
CLOSED AT 

550 

kts

kts

kts

kts

SWITCH TEST

CONTACT STATE

AIRSPEED

RUN

MANUAL 

OPERATION

STOP

MANUAL 

OPERATION

STOP

AIRSPEED

AIRSPEED

CONTACT STATE
CLOSED AT 

550 

kts

OPENED AT 

525 

kts

HYSTERESIS 25 

kts

SWITCH TEST

SWITCH TEST

CONTACT STATE
CLOSED AT 

550 

kts

Summary of Contents for Druck DPI 610A

Page 1: ...Druck com DPI 610A Aeronautical Portable Pressure Calibrator Instruction Manual...

Page 2: ......

Page 3: ...acturer s service departments Druck com For technical advice contact the manufacturer Symbols WARNING Before operating this instrument read the safety instructions Do not apply pressures greater than...

Page 4: ...18 ft 30 000 ft 36 ft Precision includes linearity hysteresis and repeatability 12 months measurement stability Temperature Effects 0 004 of reading C averaged over 10 to 40 C w r t 20 C 0 002 of read...

Page 5: ...g absolute version of the DPI 610 provides manual generation of pressure and vacuum for testing of aircraft Pitot and Static systems It has a special pneumatic manifold and output port assembly the re...

Page 6: ...ator Controls 1 Output port AN4 to pitot or static system 2 Manual let down valve 3 Electrical switch inputs 4 Release valve releases system pressure vacuum at a controlled rate 5 Vent port 6 Selects...

Page 7: ...imum measurement ratings of the standard instrument that should not be exceeded Pressure over range 120 of FS is indicated by a flashing pressure display Voltage or current over range 110 of max cause...

Page 8: ...n POD by selecting POD ADJUST as follows Apply QFE pressure reading as POD correction as follows WARNING Observe safety precautions stated in local orders and the aircraft or equipment servicing proce...

Page 9: ...RUN key F1 and operate the hand pump continue until the switch operates mbar POD Off POD Correction mbar POD Correction POD Off WARNING VENT SYSTEM BEFORE SETTING POD WARNING VENT SYSTEM BEFORE SETTIN...

Page 10: ...m is vented to atmosphere before starting the test 2 Select negative pressure and close the release valve 3 Close the manual let down valve 4 Press the RUN key F1 and operate the hand pump continue un...

Page 11: ...a new system altitude and press RUN F2 again 7 On completion of the test allow the altitude to decrease until it is below 1500 ft 8 Slowly open the manual let down valve to return the system to ground...

Page 12: ...be potential free Figure 4 Airspeed Switch Test Connections 2 Ensure that positive pressure is selected and close the release valve 3 Close the manual let down valve 4 Zero the airspeed reading by pre...

Page 13: ...ng the test 2 Ensure that positive pressure is selected and close the release valve 3 Close the manual let down valve 4 Zero the airspeed reading by pressing the ZERO key 5 Press the RUN key F1 and op...

Page 14: ...layed operating data 6 Allow the airspeed to decrease until it is below 80 kts 7 Slowly open the manual let down valve to return the system to zero ground level SELECT TASK SELECT TASK ALTITUDE SWITCH...

Page 15: ...1998 Baker Hughes Company English DPI610A Instruction Manual 13 8 To perform another leak test use the hand pump to generate a new airspeed and press RUN F2 again 9 To exit or cancel leak test mode p...

Page 16: ...Copyright 1998 Baker Hughes Company 14 DPI610A Instruction Manual English...

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ...kerhughes com Bangalore 91 9986024426 aneesh madhav bakerhughes com Milan 39 02 36 04 28 42 csd italia bakerhughes com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 3...

Reviews: