MID_PrimoAdvanced_Bedienungsanleitung_0704_e.doc
Electrical connection
Page 12/37
4.3 Terminal wiring plan of inputs and outputs
In- / Output
Description
Terminal
Analog
0 - 20 mA
4 - 20 mA
RL < 800 Ohm
0 - 10 mA
2 - 10 mA
3 (+)
1 (-)
Pulse
Active
Passive
0 - 10 kHz, pulse-/break relation approx. 1:1
adjustable pulse length 5 - 500 ms (5 ms steps)
Pulse output invertable
24 V DC, 25 mA
max. 30 V DC, 250 mA
13 (+) and 16 (-)
15 (+) and 16 (-)
Frequency
(passive)
50 - 5000 Hz
Full scale 500 - 5000 Hz
max. 30 V DC, 250 mA
6 (+)
5 (-)
Relay 1
Flow direction
max. 48 V, 500 mA
10, 11 and 12
Relay 2
Set point, preselection quantity
max. 48 V, 500 mA
8 and 7
Relay 3
Meter disturbance, exceeding of measuring range and
medium control
max. 48 V, 500 mA
8 and 9
Reset
Totalizer 2 or preselection quantity
14 and 16
RS 232
Output of current readings and parameter setting