Parameter setting
Page 14/37
Value with negative sign means flow simulation opposite to the main flow direction.
The respective relay is activated simultaneous to the chosen flow rate (see chart).
T e s t
S i g n a l
o u t
S i g n a l
=
0 %
^ h i g h e r
>
l o w e r
E C o n t i n u e
In this case „signal = 100%“ means the simulation of a flow rate of 100% of the set full
scale. By pressing the key
↑
and
→
the next resp. previous value can be chosen. This
menu point can be left with the key
E
.
5.1.3
Calibrating the analog output
The power output can be calibrated with this menu point.
Notice:
A calibration of the analog output is already done in the factory. A
recalibration is normally not necessary.
A n a l o g
o u t p u t
^ C l e a r
> C a l i b r a t i o n
E C o n t i n u e
By pressing the key
↑
the present calibration is erased, the calibration menu is shown
with the key
→
and the calibration is kept with the key
E
.
Offset (4 mA) and range (20 mA) can be set in the calibration menu. By pressing the
keys
↑
(increase) and
→
(reduce) the value can be calibrated at the analog output
with the help of a current meter. The values are acknowledged with the key E. The
calibration is also effective for the range 0-20 mA.
1. Connect a current meter at the analog output.
2. Setting of the offset (4 mA) with the
↑→
keys and acknowledgement with
E
.
3. Setting of the range (20 mA) with the
↑→
keys and acknowledgement with
E
.
5.1.4 Calibrating the medium control
Notice:
In order to compensate different conductivity of the media, cable
length or sizes at the medium control, a calibration is
recommended. This is important, if the medium control will be
activated.
E m p t y
p i p e
d e t.
^ C l e a r
> C a l i b r a t i o n
E C o n t i n u e
By pressing the key
↑
the present calibration is erased, with the key
→
the calibration
menu is shown and with the key
E
the calibration is kept.