Electronic register
page 9/11
IND_MNMR100_Bedienungsanleitung_0703_e.doc
45 LIGHT = LitE 45 N/A (turn back light on or off)
Note: The back light is not used with this model meter
46 TIME = tiME 46 N/A (set back light time)
Note: The back light is not in use with this model meter
4.6 Calibration
CALI5
Note: Provides calibration adjustment up to +/- 5% of reading. You will need to calcu-
late the percentage adjustment you require. If the meter is reading over you will
need to enter a minus figure. If the meter is reading under you will need to enter
a plus figure.
Step 1: Reset the batch reading to zero.
Step 2: Dispence at least 5 liters, using a constant flow rate, into a graduated calibrated
container.
Step 3: Take the reading from the IM50.
Step 4: Calculate the % error as follows. % error = ((container volume - IM50 volume) /
container volume) * 100.
Step 5: Enter the % error into the meter as calculated, if the value is negative enter a
negative number.
51 CALIBRATION = CALI 51 (adjust calib/- 5%)
a) Press and hold down RESET for 5 seconds until “MEtr 1” appears.
b) Press the DISP button 4 times to reach “CALI 5”
c) Press RESET “CALI 51” will appear.
d) Press and hold down RESET for 2 seconds (program will appear and the first digit
will flash)
e) *Press RESET to change number. *Press DISP to change the decimal point
position, or go to the next number. *Press the RESET button to set the negative
value.
f) Press and hold down RESET until “program” disappears to accept setting.
g) Press RESET for five seconds to return the meter to the operating mode.
4.7 Configuration
6 CONFIGURATION = CONF 6 (configuration details) CONF6
a) Press and hold down RESET for 5 seconds until “MEtr 1” appears.
b) Press the DISP button 5 times to reach “CONF 6”.
c) Press RESET “MOD” plus model number will appear.
d) Press RESET “REL” plus hardwear release number will appear.
e) Press RESET the screen will flash between “SOFT” and the softwear release no’s.
(at 2 second intervals)
f) Press RESET the screen will flash between “SEr” and the 6 figure serial number.
(at 2 second intervals)
g) Press RESET “MAN” plus manufacturer number.
h) Press RESET the screen will flash between “CONFIG” and the configuration
number (at 2 second intervals).
i) Press RESET for five seconds to return the meter to the operating mode.