Page 30
4-11
Model M-2000 Mag Meter
Meter Setup
Detector Factor
[Factory Set]
A
This parameter is set at the factory. This factor compensates for accuracy error as a result of the installed
detector. If accuracy adjustment of the meter is required, please refer to the scale factor.
In the event the amplifier is replaced, this parameter must be reprogrammed with the original detector
factor.
Detector Offset
[Factory Set]
A
This parameter is set at the factory. This parameter compensates for accuracy error as a result of the
installed detector. If accuracy adjustment of the meter is required, please refer to the scale factor.
Measurements
Flow Units
[GPM]
U
Flow Units lets you set the unit of measure for the flow rate and full scale flow. Changing the flow units
parameter readjusts the full scale flow parameter. For example, changing from GPM to GPS would
change the full scale flow from 60 GPM to 1 GPS.
To change the Flow Unit, follow these steps from the
Measurements menu
:
1. Select
Flow Units
to view the Flow Units display.
2. Position the arrow next to one of the following flow unit options:
Code
Flow Unit
Code
Flow Unit
LPS
Liter/Second
GPM
Gallons/Min.
LPM
Liter/Minute
GPH
Gallons/Hour
LPH
Liter/Hour
MGD
MegaGallon/Day
M3S
Cubic Meter/Sec.
IGS
UKG/Sec.
M3M
Cubic Meter/Min.
IGM
UKG/Min.
M3H
Cubic Meter/Hour
IGH
UKG/Hour
F3S
Cubic Feet/Sec.
LbM
Pound/Min.
F3M
Cubic Feet/Min.
OPM
Ounce/Min
F3H
Cubic Feet/Hour.
BPM
Barrel/Min
GPS
Gallons/Sec.
3. Press
[E]
to save the flow units and return to the Measurements menu.