Using the Programmer Software
Change Current Settings
To change the encoder settings and set a new default value for any of the data fields, follow this procedure .
1 . Click the drop-down menu to the right of the
field to display a list of values .
2 . Click the value you want to set as the default
for that field .
OTEE:
N
The field label changes to bold text, indicating
a new, unprogrammed value is selected . See
OTEE:
N
You must set the default value for the
Meter Type
field before you set the
Meter Model / Size
. The
Unit of Measure
value
does
not
have to be the same as the
Rate of Flow
Units.
(Even in a cubic foot meter,
the flow rate can be gallons per minute .)
Figure 13: Unprogrammed values
Apply Current Settings
If you make a change on the Programmer screen, you must apply the change to the encoder .
With the IR head aligned, click Apply Current Settings to program the new values to the encoder .
The new values display on the
Programmer
screen and the field labels are no longer bold . The encoder also goes into storage
mode . In storage mode, the encoder displays the
meter type, digit resolution from the encoder
and
unit of measure
. This is an
example of an encoder display of a Model 25 Disc Meter with a 9-digit output, measuring in gallons:
GAL
Figure 14: Meter type, digit resolution from the encoder and unit of measure
Updating Multiple Encoders
If you have multiple encoders to program with the same settings, attach the IR head bracket with the IR cable to each
encoder, one at a time, and click Apply Current Settings to program each encoder with the same settings .
Page 10
May 2018
ENC-PM-02040-EN-03
Summary of Contents for HR-E LCD 4-20
Page 2: ...High Resolution LCD Encoder Programmer Page ii May 2018 ENC PM 02040 EN 03...
Page 4: ...High Resolution LCD Encoder Programmer Page iv May 2018 ENC PM 02040 EN 03...
Page 19: ...Appendix APPENDIX Page 19 May 2018 ENC PM 02040 EN 03...
Page 23: ...Programmer Manual INTENTIONAL BLANK PAGE Page 23 May 2018 ENC PM 02040 EN 03...