Programming Procedures
Password (PASSWORD)
Password protection prevents unauthorized users from changing programming
information Initially, the password is set to all zeros Edit with the
Restore Defaults (RES DFLT)
Use this feature to restore factory calibration data To restore factory calibration data, select
YES
, then press
ENTER
Advanced Programming Descriptions
Advanced programming allows for re-configuring the analog output Calibration of the
analog output is preset at the factory, but can be changed to customize calibration for
your installation
To access the
Advanced Programming
options, press and hold
MENU
for approximately
3 seconds until "DISPLAY" shows on the display panel The programming menus begin with
the
DISPLAY mode
and continue as described above through
Output Mode
(
OUT MODE)
After you enter
Output Mode
,
Advanced Programming
starts with the following:
Calibrating Analog Output (CAL OUT?)
OTEE:
N
Setup prompts and descriptors for configuring and calibrating the analog output correspond
to the output mode selected Refer to the flowcharts on pages
through
To test or change the analog output calibration:
1 At the
CAL OUT?
prompt press
ENTER
"NO" displays
2 Press either arrow key to select
YES
3 The analog output goes to its minimum output level A numeric value between
0…4000 displays This is an internal number used to drive the analog output
4 To increase the analog output signal level, press
ñ
To decrease the analog output
signal level, press
ð
5 Press
ENTER
to store the setting
6 The analog output goes to its maximum output level A numeric value between
0…4000 displays This is an internal number used to drive the analog output
7 To increase the analog output signal level, press
ñ
To decrease the analog output
signal level, press
ð
8 Press
ENTER
to store the setting
9 The unit advances to the analog output test mode The analog output goes to its
minimum output level A numeric value of 0 displays For test purposes, the analog
output signal can be run up or down in increments of 1 milliamp or 1 volt, depending
on the
OUT MODE
selected
10 To increase the analog output signal level, press
ñ
To decrease the analog output
signal level, press
ð
11 Press
ENTER
to exit the analog calibration mode
12 The unit automatically advances to the
PASSWORD
feature
Page 19
September 2017
XMT-UM-00498-EN-03