INSTALLATION
Connecting the ER-500 Monitor to a Pulse Output Device
The ER-500 monitor has two jumpers that are used to set the type of signal and the minimum amplitude of the signal that
it accepts When used with Badger Meter IOG oval gear meters, the Input Signal Level should be set to
Low
and the Input
Waveform should be set for pulse as shown in
JP1
JP2
JP3
Input
Total P
ulse
Sig
nal
Mag
Pulse
Iso
OC
Low
High
Input Waveform Selection
(Should be set to Pulse for IOG meter)
Input Signal Level Selection
(Should be set to Low for IOG Meter)
Figure 2: Input jumper settings
If the ER-500 monitor is a replacement, it must be calibrated for the IOG it is intended to be used with The K-factor for the
specific IOG meter must be programmed into the ER-500 monitor The K-factor value is found on the calibration certificate
that came with the IOG meter For instructions on programming the K-factor, see
Enter Flow Sensor's K-factor* on page 17
.
Transmitter Connections
The ILR transmitter typically used with the IOG meter family has two sets of pulse output wires The white and green output
leads connected to the primary reed switch bank are generally the first choice (see
)
JP1
JP2
JP3
Input
Total P
ulse
Sig
nal
P1
Freq. In
4-20mA
Iso Total Pluse
TR_B
TR_A
RS485 Gnd
Setpoint 1
Setpoint 2
Gnd
+
–
+
–
+
–
Total Reset
OC Total Pluse
Signal Gnd
TB1
Mag
Pulse
Iso
OC
Low
High
Freq. In
Gnd
Mag
Pulse
+
–
Reed Switch
Bank
ILR
Pulse Transmitter
White
Green
Blue
Black
Auxiliary Reed
Switch Bank
Figure 3: Typical IOG meter input connection
The ILR transmitter also has a secondary (auxiliary) set of pulse output wires Either pair can be used to connect to the ER-500
monitor The connections are:
ILR Wires
ER-500 Terminals
Reed Switch Bank (Primary)
Reed Switch Bank (Auxiliary)
Freq In +
White
Blue
Freq In -
Green
Black
Installation
DSY-PM-00028-EN-02
Page 7
February 2017
Summary of Contents for ER-500
Page 1: ...Display ER 500 Advanced DSY PM 00028 EN 02 February 2017 Programming Manual...
Page 2: ...ER 500 Flow Monitor Page ii February 2017 DSY PM 00028 EN 02...
Page 4: ...ER 500 Flow Monitor Page iv February 2017 DSY PM 00028 EN 02...
Page 34: ...Display ER 500 Advanced INTENTIONAL BLANK PAGE DSY PM 00028 EN 02 Page 34 February 2017...
Page 35: ...Programming Manual INTENTIONAL BLANK PAGE DSY PM 00028 EN 02 Page 35 February 2017...