Linearization
To increase accuracy, linearize the monitor The linearization function accepts a maximum of ten points and requires
additional calibration data from the meter being used with the monitor Typically, calibration information can be obtained
in three, five and ten points from the flow meter’s manufacturer If linearization is not needed, press
RIGHT
to advance to
the
Modbus
parameter See
To complete linearization, press
ENTER
at the
Linear
prompt The monitor
advances to the
Lin Pts
parameter
Number of Points
The
Lin Pts
value displays If the number of points is set to
0
, linearization is disabled Press
ENTER.
The most significant digit
of the number of points entry begins to flash The first number can be a
1
or a
0
only Press
UP
to change the first digit Press
RIGHT
to move to the least significant digit
OTEE:
N
If the number you enter is out of range the, display flashes
Limit
and refuses the entry
Press
ENTER
to advance to the
Freq#1
prompt
OTEE:
N
If the number of linear points is set to 1, the ER-500 monitor assumes that you are entering the maximum frequency
and coefficient Further, the meter assumes that the implied first point is at a frequency of 0 Hz and a coefficient of 0
Frequency
At the
Freq#1
prompt, press
ENTER
The first digit of the first linear point’s frequency input flashes Press
UP
to increment
the numerical values and
RIGHT
to change the position of the number being entered When the frequency value input is
complete, press
ENTER
to save and advance to the
Coef#1
parameter
Coefficient
The coefficient is the value applied to the nominal K-factor to correct it to the exact K-factor for that point The coefficient is
calculated by dividing the average (nominal) K-factor for that point by the actual K-factor for the flow meter
Linear Coefficient =
Actual K-Factor
Nominal K-Factor
At the
Coef#1
prompt, press
ENTER
The first digit of the coefficient flashes Press
UP
to increment the digit, and
RIGHT
to
move to the next digit When all digits have been entered, press
ENTER
to save and advance to the next frequency input
Continue entering pairs of frequency and coefficient points until all data has been entered Press
ENTER
to save and advance
to the
Modbus
parameter
OTEE:
N
The frequency values must be entered in ascending order If a lower frequency value is entered after a higher value
the ER-500 monitor flashes
Limit
followed by the minimum frequency value acceptable to the display
Example:
The following is actual data taken from a one inch turbine flow sensor calibrated with water
Unit Under Test (UUT) Calibration In GPM
Actual
GPM
UUT
Frequency
UUT Actual
K-factor
(Hz x 60)
Nominal K
Linear
Coefficient
Raw Error
Hz
Counts/Gallon
GPM
% Rate
50 02
755 900
906 72
49 72
1 0060
0 59
28 12
426 000
908 96
28 02
1 0035
0 35
15 80
240 500
913 29
15 82
0 9987
-0 13
8 88
135 800
917 57
8 93
0 9941
-0 59
4 95
75 100
910 30
4 94
1 0020
0 20
Nominal K (NK)
912 144
—
—
In this example the linear coefficient has already been calculated by the calibration program so all that is required is to enter
five into the
Lin Pts
parameter and then enter, in order, the five frequency and linear coefficient data pairs
Programming
DSY-PM-00028-EN-02
Page 21
February 2017
Summary of Contents for ER-500
Page 1: ...Display ER 500 Advanced DSY PM 00028 EN 02 February 2017 Programming Manual...
Page 2: ...ER 500 Flow Monitor Page ii February 2017 DSY PM 00028 EN 02...
Page 4: ...ER 500 Flow Monitor Page iv February 2017 DSY PM 00028 EN 02...
Page 34: ...Display ER 500 Advanced INTENTIONAL BLANK PAGE DSY PM 00028 EN 02 Page 34 February 2017...
Page 35: ...Programming Manual INTENTIONAL BLANK PAGE DSY PM 00028 EN 02 Page 35 February 2017...