System Setup
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
Page
4-3
Valid Settings: Any oxygen concentration between 0 to 100%
Note: The Analog Output scale may be inverted if the voltage in User Setup 9 is set to be higher
than voltage in User Setup 10.
4.1.1.11
User Setup 11: Minimum Analog Output Current Concentration
This parameter is only valid if Menu 5 is set to option 5
– Manual Range. This sets the oxygen
concentration at which the Analog Current is at minimum output.
Valid Settings: Any oxygen concentration between 0 to 100%
4.1.1.12
User Setup 12: Analog Current Full Scale.
This parameter is only valid if Menu 5 is set to option 5
– Manual Range. This sets the oxygen
concentration at which the Analog Current is at maximum output.
Valid Settings: Any oxygen concentration between 0 to 100%
Note: The Analog Output scale may be inverted if the voltage in User Setup 11 is set to be higher
than voltage in User Setup 12.
4.1.1.13
User Setup 13: Set Current Year. ** NOT ACTIVE FOR THIS RELEASE**
4.1.1.14
User Setup 14: Set Current Month. ** NOT ACTIVE FOR THIS RELEASE**
4.1.1.15
User Setup 15: Set Current Day. ** NOT ACTIVE FOR THIS RELEASE**
4.1.1.16
User Setup 16: Set Current Hour. ** NOT ACTIVE FOR THIS RELEASE**
4.1.1.17
User Setup 17: Set Current Minute. ** NOT ACTIVE FOR THIS RELEASE**
4.1.1.18
User Setup 18: Factory Setup Restore.
This parameter allows the user to return the Mini-ICS to its initial factory-commissioned settings.
This will activate if the user adjusts the setting number
to 99, and pushes the “MODE” key. The 7-
segment alphanumeric display will flash “Fr” before the unit returns to Run mode.
Valid Settings: 99. A setting of 99 will activate the Factory Setup restore.
4.1.2 System Setup via Service Port
The model Mini-ICS features a Service Port, which is accessible for programming the system,
monitoring the analyzer output, and determining the active fault codes for troubleshooting. The
Service Port has been designed for communication with a PC based computer or other device
capable of receiving and transmitting ASCII data packets over a standard RS-232 serial interface.
Access to the Serial Service Port may made through a terminal emulator program such as
HyperTerminal, available in Microsoft Windows 95 or later:
4.1.2.1
RS-232 Service Port Interfacing with HyperTerminal in Microsoft Windows 95 or later
Turn off your PC computer, and remove power from the Model Mini-ICS. Complete the instructions
for wiring and connecting the Model Mini-ICS to a PC computer (section 2.1.3.10). Apply power to
the Model Mini-ICS, and start up the PC computer.