System Setup
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
Page
4-5
Advanced Level-2 Access: Allows access to vital control areas via password.
4.1.2.4
Standard Level Access
Standard Access level is the default level of access to the Model Mini-ICS available to the user via
a host computer or printer over a standard RS-232 serial interface. In Standard Level access, the
user can make inquiries about Oxygen concentration, sensor signal level, and other parameters for
system servicing, and troubleshooting.
When communications are established between the Model Mini-ICS and a host computer, 2-way
communication begins automatically in Standard Level access. Data is sent out the analyzer
RS232 Service Port to the host terminal once-per-second, in the factory-default human-readable
format (section 4.1.2.4.2). There are no commands necessary to begin viewing information
transmitted by the Model Mini-ICS in Standard Level access.
To request and view specific information via the RS-232 interface, type the desired command key
selected from the Standard Access level command chart below (It is not necessary to press
return).
Helpful Hint:
For viewing convenience, before requesting specific information from the Model Mini-
ICS, disable automatic 1-second updates from the Model Mini-ICS and allow access of information
by-request-only (section 4.1.2.4.1), by typing
“SSERFMT-0”, followed by the Return key. To return
to automatic 1-second updates of data from the analyzer in human-readable format (section
4.1.2.4.2), typ
e “SSERFMT-1”, followed by the Return key.
The STANDARD ACCESS level commands
TYPED COMMAND
DESCRIPTION OF QUERIED
FUNCTION
A
Short software version
B
Software Build number
C
Oxygen sensor type
E
Sensor output in Percent Oxygen
G
Sensor output in Volts or Amperes