System Setup
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
Page
4-11
4.1.2.10
(U30) Analog Output Setup Menu
The RS-232 Analog Output Setup menu U30 provides access to all of the settings related to the
Analog Voltage Output (TB3-pin5, TB3-pin6) and Analog Current Output (TB3-pin7, TB3-pin8). It
is accessed from the Setup Main Menu by typing “3” on the RS-232 terminal. To navigate
backwards, use the <Esc> or “Q” key on the RS-232 terminal.
4.1.2.10.1 Analog Voltage Output Range Jumpers
This menu sets the Analog Voltage Output full-scale value. It may be set to 0 (5V full scale), 1
(10V full scale) or 2 (1V full scale). This setting must match the RA and RB hardware jumper
settings on the bottom of the main cpu pcb (section 4.1.3). This menu is accessed from the
Analog Output Setup
menu by typing “1” on the RS-232 terminal.
4.1.2.10.2 Analog Current Output
This menu is accessed from the Analog Voltage Output Setup menu by typing “2” on the RS-232
terminal. It is used to select the analog current output to either 0 (0-20mA) or 1 (4-20mA).
4.1.2.10.3 Analog Output Calibration and Range Control
This menu is used to test and calibrate the Analog Voltage Output and Analog Current Output.
This menu also allows for the analog voltage and current ranges to be manually set to meet
application requirements
. It is accessed from the Analog Output Setup menu by typing “3” on the
RS-232 terminal.
4.1.2.10.4 Calibrate 0 To 5 Volt Output
This menu calibrates the 0-5V Analog Voltage Output. A digital multi-meter set to measure volts
must be attached to TB3-pin5 and TB3-pin6. Jumpers RA and RB must be open (section 4.1.3).
Navigate from Analog Output Setup menu to the Analog Output Calibration menu by typing “3” on
the RS-23
2 terminal. Type “2” to access the Calibrate 0 To 5 Volt Output menu.
4.1.2.10.5 Calibrate 0 To 10 Volt Output
This menu calibrates the 0-10V Analog Voltage Output. A digital multi-meter set to measure volts
must be attached to TB3-pin5 and TB3-pin6. Jumper RA must be shorted and RB must be open
(section 4.1.3).. Navigate from Analog Output Setup menu to the Analog Output Calibration menu
by typing “3” on the RS-232 terminal. Then, type “3” to access the Calibrate 0 To 10 Volt Output
menu.