Step
Bump Testing Using Calibration Gas Cylinders
1
Remove the Splash Shield accessory if one is used.
2
Connect the regulator to the test gas port using 3 mm or 1/8”
inside diameter tubing (see below).
3
Expose the gas detector to gas from the cylinder. Monitor
the LED display reading. Refer to the specifications section
for acceptance criteria.
Figure 16. Using Calibration Gas
6.4. Adjustment Using Calibration Gas
Adjustment Using Calibration Gas requires a gas cylinder with the
appropriate gas and concentration. Note that Bacharach offers a
calibration kit that consists of a calibration gas cylinder and a flow
regulation valve with flexible non-absorbent tubing.