MGS-250 Manual
6490-9000 Revision 4
55
Description
Fault Code
Hex
Code
Possible Causes
Critical Faults
Critical fault. Cycle power to
the sensor and see of the
fault clears. If not, contact
Bacharach technical support
for additional guidance.
: :
IMPORTANT:
Fault F.-15 may be caused by rapid changes
in temperature, and other environmental effects. If the F.-15
fault remains after the ambient temperature has stabilized,
perform a manual re-zero to clear the fault. See Section 4.3.
8.2. Diagnostics Attributes (P.-18)
Use this option to review the built-in diagnostic attributes. Access the
parameter list (see Section 5) and select P.-18. Press the Enter button to
access the diagnostics, and then use the UP or DOWN button to select
each attribute. The following information is available.
Table 19: Diagnostic Attributes
Attribute
Description
Displays the current fault condition code in hex format:
0000
= no faults are active
XXXX
= HEX number
See Table 18 for the HEX format cross reference.
Displays a hex format code that corresponds to any
faults that occurred since the internal fault record was
last erased. See Table 18 for the HEX format cross
reference. The records may be erased by pressing the
ENTER button.
Displays the number of days that the MGS-250 has
been in service. The value of this attribute rolls over
after 9999.