Legend
1 – on/off button
2 – numerical display
3 – two piezo speakers (acoustic alarm)
4 – upper plastic cover
5 – battery compartment (for two alkaline AA batteries)
6 – green indicator light (sensor pad status), BC-220i and BC-230i have double indicator lights, i.e. green and red combined
in one
7 – red indicator light (alarm status), BC-220i and BC-230i have double indicator lights, i.e. green and red combined in one
8 – control unit
9 – left socket (to connect sensor pads)
10 – right socket (to connect sensor pads)
11 – sensor pad wrapped in plastic
12 – T-connection (for more pads or for twins)
Putting Baby Control Digital breathing monitor to operation
(BC-200, BC-210, BC-230)
1. Detach the upper plastic cover (4) from the control unit (8). Parts are not fixed with any screw, it is only necessary to put
them apart carefully.
2. Insert two alkaline AA batteries (do not use rechargeable batteries) into the battery compartment (5) inside the control
unit (8). Make sure the batteries are in the /– orientation.
3. Subsequently, the condition of batteries and control unit (8) are automatically tested. You will hear an acoustic signal from
the speaker (3) twice and the green indicator light (6) flashes once and the red indicator light (7) flashes twice.
4. Attach the upper plastic cover (4) to the control unit (8).
5. Connect the sensor pad cable (11) to the socket (9 or 10) located on the side of the control unit (8). The sensor pad (11) can
be connected to any of the sockets (9 or 10).
6. Place the sensor pad (11) evenly at the bed’s grating approximately on the place where the baby’s chest is to be situated
and place the mattress over it. Never bend, fold or roll sensor pads!
7. Attach the control unit (8) to the baby’s bed using Velcro clamps. Baby Control Digital cable must be secured so that the
baby cannot pull it. The best way is to turnit around one of the bed bars.
8. Baby Control Digital breathing monitor is ready for operation.
INFORMATION:
When one sensor pad is used (BC-200 type), we recommend placing it under the mattress
lengthwise for safety reasons; when two and more sensor pads are used (BC-210 and BC-230 types), place them
crosswise. If all three sensor pads are needed (BC-230 type), connect them to the control unit using the T-con-
nection (part of delivery). If the baby is already able to leave the place, the use of two or three sensor pads is
advised, even though one sensor pad would fit the baby. It may occur that the baby moves off one sensor pad
during sleep and so the other pad can take over the sensing function.
Typy monitorů dechu Baby Control Digital BC-200, BC-210, BC-230 používejte pouze pro jedno dítě.
U typů BC-220i a BC-230i je možné v případě přesného nastavení (viz REŽIM DVOJČATA) rozdělení senzorových
podložek mezi dvě děti.
IMPORTANT
: In rare cases, especially with low weight babies or babies lying on their sides, it may occur that
the sensor pad senses poorly and the displayed measured value is often below the pre-set threshold level of
detection. A temporary solution may be to readjust the threshold level of detection (see Adjusting the threshold
level of detection).