background image

To ensure proper functioning of the twin mode, each baby must be in placed his/her own bed. Beds must be placed from 

each other as far as the cables of the sensor pads make it possible, but at least 40 cm. This way the movements of babies will 

not interfere with one another’s sensor pads. You may test whether the beds are in adequate distance by taking one baby 

out of the bed. The sensor pad placed in his/her bed then must display the state with no movement and Baby Control Digital 

breathing monitor must trigger an alarm. Then the test should also be carried out with the second baby. At least 2 sensor 

pads are needed for the twins, one pad under each baby. 

Putting Baby Control Digital breathing monitor – TWIN MODE 

(BC-220i, BC-230i)

1.  Detach the upper plastic cover (4) from the control unit (8). Parts are not fixed with any screw, it is only necessary to put

     them apart carefully. 

2.  Insert two alkaline AA batteries (do not use rechargeable batteries) into the battery compartment (5) inside the control

     unit (8). Make sure the batteries are in the /– orientation.

3.  Subsequently, the condition of batteries and control unit (8) are automatically tested. You will hear an acoustic signal from

      the speaker (3) twice and simultaneously the left indicator light (6) flashes green and the right indicator light (7) flashes red.

4.  Attach the upper plastic cover (4) to the control unit (8).

5.  Connect the sensor pad cables (11) to the socket (9 or 10) located on the side of the control unit (8).   

6.  Place the sensor pads (11) evenly at the bed’s grating (approximately on the place where the baby’s chest is to be situated)

     and place the mattress over them. Never bend, fold or roll sensor pads!

7.  Attach the control unit (8) to the baby bed using Velcro clamps. Baby Control Digital cable must be secures so that the

     larger baby cannot pull it, the best way is to turn it around one of the bed bars

8.  Baby Control Digital breathing monitor is ready for operation.

Switching ON/OFF Baby Control Digital breathing monitor – 

TWIN MODE   (BC-220i, BC-230i)

1.  Press and hold the on/off button (1) for 2 seconds. Baby Control Digital breathing monitor will not start by any shorter

     pressing.

2.  The pre-set threshold level of detection will flash on the display (2) ten times. The factory setting is “3“. 

3.  Then the numerical values „0” = no movement, „1” to „9” = there is any movement will repeatedly flash on the display (2).

      At first, the measured value coming from the left socket (9) is displayed with simultaneous flashing of the left indicator light

     (6), then from the right socket (10) with simultaneous flashing of the right indicator light (7). 

4.  Baby Control Digital breathing monitor is switched on and ready for operation.

5.  The indicator lights (6 and 7) flash green when the intensity of breathing movements reaches or exceeds the pre-set

     threshold level of detection.

6.  The indicator light (6 or 7) will start to flash red accompanied by a short acoustic signal from the speaker (3) if a dangerous

     situation occurs (see Alarm status of Baby Control Digital breathing monitor - TWIN MODE).

7.  After the alarm acoustic signal is activated, you can stop the alarm by only pressing the on/off button (1) once. The

     control unit (8) is switched off. 

8.  Baby Control Digital breathing monitor can be switched off by pressing the on/off button (1) twice in succession. After

     the first pressing the display (2) shows „-” and if you press the on/off button (1) again within 5 seconds, the device

     switches off with two short acoustic signals. If you do not press the on/off button again within 5 seconds, the monitor

     stays on and returns to normal operation.

Baby Control Digital   

 

 

 

 

 

 

           12

Summary of Contents for BC-200

Page 1: ...for Baby Control Digital breathing monitor BC 200 with one sensor pad BC 210 with two sensor pads BC 230 with three sensor pads BC 220i with two sensor pads may be used with twin babies BC 230i with f...

Page 2: ...ore breathing the baby is likely to be saved Using the breathing monitor is the most efficient way to prevent SIDS Baby Control Digital breathing monitor can save life but is not a life saving device...

Page 3: ...in the dark luminous Thus it is possible to check with only one look whether Baby Control Digital breathing monitor is appropriately adjusted for your baby s breathing 8 Baby Control Digital breathin...

Page 4: ...20i breathing monitor for twins package includes 1 pc BC 104 control unit usable both in one baby and twin baby mode 2 pcs BC 101 sensor pads wrapped in plastic without phthalates according to the Eur...

Page 5: ...oth environment and baby friendly Optimallistening distance is 15 30 metres The supplementary BC Wireless unit provides an excellent solution if you must stay away from your baby but wish to see about...

Page 6: ...6 Place the sensor pad 11 evenly at the bed s grating approximately on the place where the baby s chest is to be situated and place the mattress over it Never bend fold or roll sensor pads 7 Attach t...

Page 7: ...arm acoustic signal is activated you can stop the alarm by pressing the on off button 1 once The control unit 8 is switched off 8 Baby Control Digital breathing monitor can be switched off by pressing...

Page 8: ...creasing the threshold level of detection Baby Control Digital breathing monitor will become more sensitive to false vibrations from the surroundings while its increase may make it more difficult to d...

Page 9: ...reathing movements is restored in this phase Baby Control Digital continues its normal operation 3 The alarm is triggered The display 2 shows the letters H E L P 4 The alarm sounds until you press the...

Page 10: ...will switch it off 6 After the whole recording has been played its end is marked with a long tone and the device switches to normal mode Low voltage of Baby Control Digital breathing monitor battery B...

Page 11: ...Control Digital breathing monitor at least once or twice a week IMPORTANT Baby Control Digital breathing monitor is sensitive to even tiny mechanical vibrations coming from the surroundings and its o...

Page 12: ...to the baby bed using Velcro clamps Baby Control Digital cable must be secures so that the larger baby cannot pull it the best way is to turn it around one of the bed bars 8 Baby Control Digital breat...

Page 13: ...RTANT For the Baby Control Digital breathing monitor operation it is very important to adjust the threshold level of detection i e the value between 1 and 9 to be attained by the breathing movement of...

Page 14: ...rol Digital breathing monitor offers you the option of temporary suspension of the twin mode in case that one baby is still sleeping but the other had to be taken out of the bed for any reason 1 For t...

Page 15: ...52 36 20 mm Number of channels 83 Weight 20 g BC Wireless BCR 402 transmitter Control unit BC 200 BC 210 BC 230 BC 220i BC 230i Sensor pad BC 200 BC 210 BC 230 BC 220i BC 230i Receiving unit Li accumu...

Page 16: ...IEC 17050 1 2010 GYEMSZI EMKI Clinically tested Certification EU Manufacturer VMD 95 Kereskedelmi s Szolg ltat Kft Sz nj tsz u 19 H 1163 Budapest Hungary kapcsolat babycontroldigital hu 36 1 401 09 64...