To ensure proper functioning of the twin mode, each baby must be in placed his/her own bed. Beds must be placed from
each other as far as the cables of the sensor pads make it possible, but at least 40 cm. This way the movements of babies will
not interfere with one another’s sensor pads. You may test whether the beds are in adequate distance by taking one baby
out of the bed. The sensor pad placed in his/her bed then must display the state with no movement and Baby Control Digital
breathing monitor must trigger an alarm. Then the test should also be carried out with the second baby. At least 2 sensor
pads are needed for the twins, one pad under each baby.
Putting Baby Control Digital breathing monitor – TWIN MODE
(BC-220i, BC-230i)
1. Detach the upper plastic cover (4) from the control unit (8). Parts are not fixed with any screw, it is only necessary to put
them apart carefully.
2. Insert two alkaline AA batteries (do not use rechargeable batteries) into the battery compartment (5) inside the control
unit (8). Make sure the batteries are in the /– orientation.
3. Subsequently, the condition of batteries and control unit (8) are automatically tested. You will hear an acoustic signal from
the speaker (3) twice and simultaneously the left indicator light (6) flashes green and the right indicator light (7) flashes red.
4. Attach the upper plastic cover (4) to the control unit (8).
5. Connect the sensor pad cables (11) to the socket (9 or 10) located on the side of the control unit (8).
6. Place the sensor pads (11) evenly at the bed’s grating (approximately on the place where the baby’s chest is to be situated)
and place the mattress over them. Never bend, fold or roll sensor pads!
7. Attach the control unit (8) to the baby bed using Velcro clamps. Baby Control Digital cable must be secures so that the
larger baby cannot pull it, the best way is to turn it around one of the bed bars
8. Baby Control Digital breathing monitor is ready for operation.
Switching ON/OFF Baby Control Digital breathing monitor –
TWIN MODE (BC-220i, BC-230i)
1. Press and hold the on/off button (1) for 2 seconds. Baby Control Digital breathing monitor will not start by any shorter
pressing.
2. The pre-set threshold level of detection will flash on the display (2) ten times. The factory setting is “3“.
3. Then the numerical values „0” = no movement, „1” to „9” = there is any movement will repeatedly flash on the display (2).
At first, the measured value coming from the left socket (9) is displayed with simultaneous flashing of the left indicator light
(6), then from the right socket (10) with simultaneous flashing of the right indicator light (7).
4. Baby Control Digital breathing monitor is switched on and ready for operation.
5. The indicator lights (6 and 7) flash green when the intensity of breathing movements reaches or exceeds the pre-set
threshold level of detection.
6. The indicator light (6 or 7) will start to flash red accompanied by a short acoustic signal from the speaker (3) if a dangerous
situation occurs (see Alarm status of Baby Control Digital breathing monitor - TWIN MODE).
7. After the alarm acoustic signal is activated, you can stop the alarm by only pressing the on/off button (1) once. The
control unit (8) is switched off.
8. Baby Control Digital breathing monitor can be switched off by pressing the on/off button (1) twice in succession. After
the first pressing the display (2) shows „-” and if you press the on/off button (1) again within 5 seconds, the device
switches off with two short acoustic signals. If you do not press the on/off button again within 5 seconds, the monitor
stays on and returns to normal operation.