Event memory of Baby Control Digital breathing monitor
(BC-200, BC-210, BC-230, BC-220i, BC-230i)
Baby Control Digital breathing monitor features 9 minutes event memory which records the events of last 9 minutes before
the switch-off (both numbers on the display and acoustic signals). These recordings may become very useful especially when
alarm occurs as they make possible to gain information about whether your baby’s breathing pattern before the alarm was
regular or not.
How to read the event memory
1. Switch off Baby Control Digital breathing monitor using on/off button (1) and wait for 5 seconds.
2. Then press the on/off button (1) and hold for 10 seconds until letters „P” - „L” (PLAY) appear. Release the on/off button (1).
3. The display (2) shows the minutes passed „1P”, „2P” to „9P” (max. 9 minutes). Choose how many minutes you want to
replay.
4. When you see the required minute, press the on/off button (1). The values, which have been recorded during this minute,
will be replayed (both numbers on the display and acoustic signals). Replay takes place in real time, which means that
recorded values follow each other in the same way and same time as they happened while the device had been used.
5. You may stop replaying at any time by pressing the on/off button (1). Pressing the button once will switch Baby Control
Digital to normal mode, pressing it twice will switch it off.
6. After the whole recording has been played, its end is marked with a long tone and the device switches to normal mode.
Low voltage of Baby Control Digital breathing monitor battery
(BC-200, BC-210, BC-230, BC-220i, BC-230i)
After Baby Control Digital is switched on by the on/off button (1) and after the pre-set threshold level of detection flashes
five times on the display (2), the letter „E” will be constantly flashing on the display accompanied by an acoustic signal. This
situation means that the battery voltage is low. Any voltage drop to 2.7V (1.35V per one battery) may endanger the reliable
operation of Baby Control Digital breathing monitor. In such case do not use the device for the reasons of safety. Baby Control
Digital cannot be switched off until batteries are replaced. Always keep spare alkaline AA batteries at home in order to ensure
uninterrupted operation. Battery life expectancy is about 2-3 months, depending on the battery type and number of hours
of operation.
IMPORTANT:
If there is a short acoustic signal after the device has been switched on and
„E” flashes once on
the display, the battery will soon get exhausted.
Baby Control Digital breathing monitor may be safely used for
another 12 hours, however you should
replace the batteries as soon as possible
!
Baby Control Digital breathing monitor must not be used with rechargeable AA batteries, because their com-
bined rated voltage is only 2.4V!
The battery condition is tested automatically after every start of the device.
Baby Control Digital