TP 10 User’s manual
20
A. Battery Pack Specifications (optional kit)
Battery Model
BP-WMP226 22/3900 SA
Battery Type
Li-ion 2S2P
Minimum Capacity
3900 mAh
Nominal Voltage
7.2 V
Max. Charge Voltage
8.4V
Cut Off Voltage
6.0v
Suggested Charge
Current (Max.)
2A
System Continuous
Discharging Current
(Max.)
16.6 A
The End of Charge
Condition
150 mA/min
Discharge Protection
UVP/OCP
Charge Protection
OVP/OTP
Self-discharge Rate
10uA ~800 uA
Dimensions
133 x 47 x 21mm
Weight
220g+/-20g
Ambient Temperature
0°C ~ +45°C
Storage Temperature
-20°C ~ +60°C
Energy
28.08Wh
Backup
100 W/ 10 min
Alarm!
Battery mode only support 100W.
If System power consumption more than 100W, Adapter mode
change to Battery mode, there is a chance to will shutdown.