TP 10 User’s manual
8
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distanced
in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Guidance
and manufacturer’s declaration – electromagnetic
immunity
The model TP 10 is intended for use in the electromagnetic environment
specified below. The customer or the user of the model TP 10 should
assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
–
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±
8 kV contact
±
15 kV air
±
8 kV contact
±
15 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least 30
%.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV for
power
supply lines
±
1 kV for
input/output
lines
±
2 kV for
power
supply lines
±
1 kV for
input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±
1 kV line(s)
to line(s)
±
2 kV line(s)
±
1 kV line(s)
to line(s)
±
2 kV line(s)
Mains power quality
should be that of a
typical commercial or
hospital environment.
Summary of Contents for e-medic Silence TP 10
Page 2: ...ii Version Change History Date Version Description Remark 2022 10 19 1 0 First Release...
Page 5: ...v Ne modifiez pas cet quipement sans l autorisation du fabricant...
Page 28: ...TP 10 User s manual 14 Dimension TP 10 VESA Mount Screw type M4...
Page 32: ...TP 10 User s manual 18 I O parts Standard version I O parts Isolated version...
Page 36: ...TP 10 User s manual 22 The panel is only intended to be used with SIP SOP facing downward...