4 Design and Construction
14
DHV0053 b2 EN
©
b2electronic
gmbh
Switches and Controls
Pos. Name
Description
1
Emergency OFF
Activates emergency shutdown.
Operation is only possible when Emergency OFF is deactivated.
•
To Activate Emergency Off
Press in
•
To Deactivate Emergency Off
Release latch and rotate
2
LED green
Indicates HV status.
Green light indicates
High Voltage is OFF
3
LED red
Indicates HV status.
Red light indicates
High Voltage is ON (possible DANGER)
DUT not discharged
(residual voltage >100V)
4
Display screen
Displays menu, options and status information.
5
Navigation knob
Enables user to select options and functions shown on display.
See 4.2 User Interface
•
To scroll selection up or down
Rotate
•
To enter selection
Click (push in)
6
HV switch [on/off]
Activates high voltage.
•
To activate HV output
Press within 10 seconds after START
See 5.2 Manual Test Procedure
See 5.3 Automatic Test Procedure