6 Tangens Delta
40
DHV0053 b2 EN
©
b2electronic
gmbh
Step
Procedure
S3
Connect HV Test Lead
•
Screw the HV test lead into the HVA HV output connector (11)
•
Connect the HV cable shield to earth.
•
Connect other end of HV test lead (clamp including screen protector) to the
DUT.
S4
For 90% it is sufficient to clean the terminations. If necessary connect the
external guard like it’s shown in the pictures below (GH0584).
S5
Verify Connections
•
Check that all cables are attached securely.
S6
Configure communication port (13)
S7
Turn key switch (7) to the “ON” position
S8
The HVA system automatically boots.
•
Start-up default screen appears
”Start Test” screen
Select appropriate option from default screen and proceed to appropriate section
for further instructions:
•
See 5.2 Manual Test Mode or
See 5.3 Automatic Test Mode
Guard
GH0584
External Guard