25
changes in to memory and add a new step in List No. 0. The page with Empty Step field in
yellow will change to 97 and Current Step to 3. All of the steps 0, 1 and 2 have been
entered.
Press the ESC key to exit. Make sure Set Repeat is = to 1 and press On/Off key to activate
the List mode.
Figure 23 - Oscilloscope view of List mode waveform example
7.2.3
Pulse Mode
The following example will describe how to setup the Pulse Mode.
Parameters
Range
Description
Volts (rms)
0 to 300.0 Vrms
Set the AC voltage of a pulse
Volts (DC)
-424.0 to 424.0 V
Set the DC voltage of a pulse
Freq.
43 to 1200.0Hz
Set the frequency of a pulse
Duty
0 to 100.0 %
Set the duty cycle for the pulse
holding (% of period time)
Period
0 to 100000ms
Set the pulse to pulse period
Count
1 to 99
Set how many pulses to generate
Sync Source
Immed, Phase
Select the mode of output
transient phase
Phase
0.0
゜
to 359.7
゜
Set pulse phase angle
Table 9 - Pulse Mode
ParametersParameters
Settings
Volts (rms)
100
Freq.
50
Duty (%)
25
Period(ms)
80
Count
4
Sync Source
Phase
Phase (deg)
90
Table 10 – Pulse Mode Settings
Summary of Contents for 9830B Series
Page 1: ...1 9830B Series USER MANUAL Model 9832B 9833B High Power Programmable AC Power Sources ...
Page 10: ...ix 2 Notations TEXT Denotes a softkey TEXT Denotes a front panel key ...
Page 28: ...15 6 Menu Tree ...
Page 86: ...73 THD 11 Harmonic 3 weight 21 3 THD 12 Harmonic 3 weight 24 5 ...
Page 88: ...75 THD 15 Harmonic 5 7 weight 2 45 3 95 THD 16 Harmonic 3 5 7 9 weight 11 4 05 2 1 3 ...
Page 89: ...76 THD 17 Harmonic 3 5 9 weight 7 17 3 42 0 8 THD 18 Harmonic 3 5 9 weight 8 11 3 48 1 ...
Page 95: ...82 THD 29 Harmonic 3 5 7 9 11 13 15 weight 33 3 20 13 8 10 8 8 5 7 2 5 5 ...