30
Operation Under Extreme Conditions
Ambient Temperature in the range of +5 to +50
o
C
Between 5 and 40 degrees functioning of the machine should not be affected.
Extremes of temperature (below 5
o
C, above 40
o
C) may affect the colour of the LCD
display. This will return to normal with the temperature.
Operation above 40 degrees is not recommended. The machine may overheat at
elevated temperatures. A visual alarm will be activated in the event of over
temperature (Internal ambient >50). Air conditioning should be employed to keep the
room temperature below 40 degrees.
Ambient Relative Humidity in the range of 10 to 100% RH
The machine is expected to function correctly at extremes of humidity. High humidity
levels may affect the colour of the LCD display. This will return to normal with the
humidity.
Atmospheric Pressure in the range of 600mBar to 1100mBar
The machine is expected to function correctly between 600 and 1100 mBar.
Supply Voltage Range from –20% to +10% of specified value
The Nippy Clearway will operate normally.
Failure of Electrical Power Supply
The machine may be operated normally using an external battery.
During total power failure,
there will be no output from the machine. The patient
will be able to breathe spontaneously through the machine and out through the
exhale port. However, some re-breathing of exhaled gas is inevitable. During
power/machine failure disconnect the patient from the breathing circuit as soon
as possible.
The inspiratory / expiratory resistance of the Nippy Clearway and breathing system
(Nippy Clearway and circuit) is less than 6cm H
2
O @ 60 l/min. This value must not
be exceeded when adding attachments or fittings to the breathing circuit.
Accessories and Spares
1.
A range of mouth pieces and facemasks is available in various sizes.
Please contact us for details
2.
A range of breathing circuits is available for use with mouth piece, facemask
or tracheotomy.
3.
Input Air Filter Element pt.no. 0960.116 (pack of 5)
4.
Inline Bacterial Filter pt.no. 0960.107 - 99.999% filtration – Resistance,
0.75mB @ 50 l/min – deadspace 55ml – 22mm tapered fittings.
These components listed above are for single patient use.
5.
External battery, part number 0960.111 (0998)
6.
Hand held switch. Part number 0960.109