24
NIV Mode
A few minutes non-invasive ventilation may be used to help the patient recover from
the exertion of the cough assist treatment. The Clearway will cycle between IPAP
and EPAP at the back up rate.
Press the
Mode
switch. Select
NIV
from the mode select screen.
Set the desired IPAP, Ti, EPAP, and
Rate.
Press the
Start/Stop
switch to begin.
Press again to stop.
Ins trigger sensitivity may be adjusted
by selecting Trigger levels from the
main menu.
NIV will continue for a maximum of 15 minutes. Do not leave the patient
unattended.
At the end of treatment a warning will be displayed on screen
accompanied by a double beep sound.
Manual override is NOT available in NIV mode.
Setting up the Clearway in NIV mode.
•
Switch on the clearway and select NIV mode from the Mode select list. (If NIV
does not appear in the list, this function is not available).
•
Press INS to select the inspiratory pressure (IPAP) and use the
◄
- and
+
►
buttons to adjust to the required pressure.
•
Press Ti to select the inspiratory time (Ti) and use the
◄
- and +
►
buttons to
adjust to the required time.
•
Press INS Rpt. to select the back up rate and use the
◄
- and +
►
buttons to
adjust to the required rate.
•
Press Exs to select the expiratory pressure (EPAP) and use the
◄
- and
+
►
buttons to adjust to the required pressure.
•
Press menu and select treatment review screen to view the waveform if desired.
Press Menu again to escape.
•
Press Start/Stop to begin treatment.