25
Advanced Settings
Vibration of airway pressure
In some cases it may be benificial in the mobilisation of secretions to “vibrate” the
airway pressure. The Nippy Clearway can be set to vibrate the insufflation or
exsufflation pressure at 4 cycles per second. The level of vibration can be preset to
25%, 50%, 75% or 100% of the working pressure.
The Airway vibration function may be accessed
via the
Advanced Settings
screen.
Press
Menu
and select
Settings and Options
and then
Advanced Settings
from the list.
Use the
Set
button to move between Ins and
Exs vibration and post breaths settings.
Use the
◄
- and +
►
buttons to select the desired
vibration level and then
Set
to accept.
Default setting is for vibration settings to be off.
Post exsufflation breaths
The post exsufflation breaths function may be accessed via the
Advanced Settings
screen.
Press
Menu
and select
Settings and Options
and then
Advanced Settings
from the
list.
Use the
Set
button to move between Ins and Exs vibration and post breaths settings.
Use the
◄
- and +
►
buttons to toggle between on and off. Press
Set
to accept.
Default setting is for post exsufflation breaths to be off.
Compliance data
Up to 28 days compliance data may be viewed on the compliance data screen.
Press Menu and select Compliance Data from
the list.
The compliance data is displayed in a
chronological list format.
Use the
◄
- and +
►
buttons to scroll up and
down the list.