29
Connecting auxiliary monitoring equipment
For monitoring or downloading data the Nippy Clearway may be connected to a
PC or Laptop computer. The Nippy Clearway isolated RS232 port is safe for use
with any domestic PC or laptop computer. However, when assembling a system,
the completed system should comply with EN60601-1 (medical systems). For
example, most computers do not comply with this standard, so it should be sited
at a distance, which makes it impossible to touch the computer and the patient at
the same time.
1. Fresh air inlet
2. Vacuum connection
3. Pressure Connection
4. Outlet connector
5.
6.
7. Pressure transducer 2
8. Exhaust Port
9.
10. Pressure / Exhalation Port
11.
Mask Pressure port
Pressure transducer 1
Breathing Circuit
Patient Connection Port
Clearway System Pneumatic Diagram
2
3
4
5
6
7
11
9
Blower
Valve
10
8
Summary of Contents for NIPPY Clearway