Apex Locator Troubleshooting

 

This guide is suitable for the users who use B.A. International’s apex 

 

locator for the first time, and those who get unstable readout because of 

 

improper operation.

Problems

Possible Causes

Analysis

Solutions

No readouts 

or unstable 

readouts

File clip

File clip wire broken or poor contact

Replace the file clip

Measuring wire

Measuring wire broken or poor contact

Poor contact between measuring wire and 

socket

Replace the measuring wire

Make sure of proper contact between socket and measuring wire

Root canal problem

Endo file is too small for a large root canal

The root canal is rinsed by pure water

Root canal is blocked by dentin chippings or 

residual pulp

Root canal is blocked by gum 

Root canal is too dry

Use the endo file with larger diameter 

Rinse root canal with saline solution

Remove the dentin chippings or residual pulp

Remove the gum

Moist the root canal with saline solution.

Metal prosthesis

Metal prosthesis

Avoid endo file touching metal prosthesis

Lip hook

Poor contact between lip hook and measuring 

wire socket

Reconnect the lip hook or replace lip hook

The screen 

displays that 

endo file reaches 

apical foramen 

before the endo 

file actually does

Endo file

Endo file hasn't actually got into root canal

Keep pushing endo file near to apical foramen, screen display will become normal

Metal prosthesis

Flush fluid touches the metal prosthesis of dental 

crown

Endo file touches metal prosthesis

Reduce flush fluid, avoid contacting metal prosthesis

Avoid endo file touching metal prosthesis

Dental crown 

problem

Blood oozes out to dental crown

Dental crown is broken or gum hyperplasia

Completely stop bleeding and clean the blood

Insulate the endo file and dental crown with insulator or cut off the gum hyperplasia

Pulp cavity problem

Root canal is cracked

There are residue, metal residue or debri in pulp 

cavity

Combine X-ray film to get the length

Remove residue, metal residue and debris

Perforation

Endodontic perforation

Repair perforation

Cavity problem

Proximal caries

Clean necrotic tissue of the proximal caries, if it's the same with adjacent dental 

surface and periodontium, temporarily fill the adjacent surface

Handing method 

shows that endo 

file reaches 

apical foramen, 

but apex locator 

shows it hasn't

File clip

Poor contact in file clip wire

Replace file clip

Root canal problem

The root canal is rinsed by pure water 

Root canal is too dry

There's ledge in root canal(without X ray film)

There are gums and dental debris

Rinse root canal with saline solution 

Moist the root canal with saline solution

With help of X ray film, avoid touching ledge with endo file

Remove the gutta-percha or debris and moist the root canal

Summary of Contents for BA OPTIMA E+ BAE380R

Page 1: ...BA OPTIMA E BAE380R Endodontic Motor INSTRUCTIONS FOR USE Please read this manual before operating BA Code BA182380 BAE380R195 ...

Page 2: ...0 degrees rotation of contra angle d Adopt real time feedback technology and dynamic torque control effectively preventing file separation 1 3 Model and specification Optima E BAE380R Endodontic Motor Please see section 1 9 1 10 for device specification 1 4 Device parts and accessories The device is composed of charging base motor handpiece contra angle Additional accessories included in the box a...

Page 3: ...precautions regarding electromagnetic compatibility EMC and must be in strict accordance with the EMC information for installation and use Do not use this equipment especially in the vicinity of fluorescent lamps radio transmitting devices remote control devices handheld and mobile high frequency communication devices 1 7 5 Please use the original contra angle Otherwise the device will be unusable...

Page 4: ...y 1 8 3 Degree of protection against electric shock B type applied part 1 8 4 Degree of protection against harmful ingress of water Ordinary equipment IPX0 1 8 5 Degree of safety application in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide Equipment cannot be used in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 1 8 6 Applied part...

Page 5: ... 2 2 1 4 SGP Mode Safety Glide Path Mode F Forward angle R Reverse angle The rotation angle is adjustable but the forward angle must be equal to the reverse angle 2 2 1 5 ATR Mode ATR Adaptive Torque Reverse function Normal continuous forward rotation when the load of the file is greater than the set torque limit the file will start to rotate forward reverse alternately at the set angle 2 2 2 Torq...

Page 6: ...device Otherwise the contra angle will be damaged 2 4 Installation and removal of contra angle 2 4 1 Installation Align any locating pin of the contra angle with the positioning slot on the motor handpiece and push the contra angle horizontally The three locating pins on the contra angle are inserted into the three positioning holes on the motor handpiece A click sound indicates that the installat...

Page 7: ...fore inserting and pulling out the file the motor handpiece must be stopped Be careful when removing files to avoid injury to fingers Removing files without holding the push button will damage the chuck of contra angle 2 6 Canal measurement functional connection This is not required if the canal measurement function will not be used Connect the measuring wire to the motor handpiece Line up the mea...

Page 8: ... isolation sleeves 2 7 1 Installation Before each use of the handpiece and after the handpiece is cleaned and disinfected put on a disposable isolation sleeve Take the isolation sleeve out of the isolation sleeve box then insert the isolation sleeve into the motor handpiece from the thin end of the handpiece and install the isolation sleeve until there is no obvious wrinkle After installing the di...

Page 9: ...tand alone apex locator AP Apical foramen Apical Action The file action when file tip reaches the flash bar point Flash Bar Position Shows the point inside the canal where specified apical action is triggered Auto Start The file rotation starts automatically when the file is inserted in the canal Auto Stop The file rotation stops automatically when the file is taken out of the canal Apical Slow Do...

Page 10: ...de and other medicant When this mode is being used a double beep sounds continuously used for indicating counter clockwise rotation happening Repeatedly press Setting button P to check that all the next level parameters of this operation mode are correct press Adjusting button to make changes if not The speed setting can be adjusted from 100 rpm to 2500 rpm Press Adjusting button to increase or de...

Page 11: ...ogical apical foramen The reference point flash bar can be set from 2 to AP Apex on the meter Rotation automatically slows down as the file tip approaches the reference point P ress Adjusting button to change OFF Disable Apical Slow Down function ON Rotation automatically slows down as the file tip approaches the reference point Forward Angle In the SGP mode the Forward Angles of 20 400 are availa...

Page 12: ...1 minute increments Press Setting button P again the Auto Standby Scr can be changed press Adjusting button to adjust then press to Main button to confirm This is the automatic return to standby display of motor handpiece when no buttons are pressed It can be set from 3 to 30 seconds in 1 second increments Press Setting button P again the Dominant Hand can be changed press Adjusting button to adju...

Page 13: ... please see chapter 3 2 Screen display Motor combined with canal measurement function mode When using motor combined with canal measurement function the measuring wire must be connected with motor handpiece by USB socket and white socket connects with patient s lip by lip hook keep the black socket idle The canal length indicator bar will show on the screen more information about canal length indi...

Page 14: ...ery time before use Clip the holder onto lip hook and check that all the bars on the meter on the screen light up otherwise the measuring wire or file clip should be replaced Root canals not suitable for canal measurement Accurate measurement cannot be obtained for the root canal conditions shown below Root canal with a large apical foramen Root canal that has an exceptionally large apical foramen...

Page 15: ...ing result between apex locator reading and radiography Sometimes the reading of apex locator and the X ray image will not correspond This does not mean that the apex locator is not working properly or that the X ray exposure is a failure An X ray image might not show the apex correctly depending on the angle of the X ray beam and the location of the apex might seem to be other than it really is T...

Page 16: ... measurement Change to CW CCW SGP or ATR mode There is continuous beep sounds after starting the motor handpiece The continuous beep sound is indicating that the motor handpiece is under CCW mode Stop the motor handpiece and change the operating mode to CW Mode Contra angle calibration failure Calibration failure caused by strong resistance of contra angle Clean the contra angle and recalibrate af...

Page 17: ...ce of the component with a dry soft nap free cloth 4 Repeat the above steps at least 3 times Note a Use distilled water or deionized water for cleaning at room temperature 6 3 1 2 Manual disinfection steps 1 Soak the dry soft cloth with 75 alcohol 2 Wipe all surfaces of headpiece charger base and other components with a wet soft cloth for at least 3 minutes 3 Wipe the surface of the component with...

Page 18: ...atment The post operative treatment must be carried out immediately no later than 30 minutes after the completion of the operation The steps are as follows 1 Remove the products from the base and rinse away the dirt on the surface of handpiece with pure water or distilled water deionized water 2 Dry the products with a clean soft cloth and place it in a clean tray Notes a The water used here must ...

Page 19: ...provided by the equipment manufacturer to familiarize yourself with the disinfection process and precautions b With this equipment cleaning disinfection and drying will be carried out together c Cleaning c1 The cleaning procedure should be suitable for the product to be treated The flushing period should be sufficient 5 10 minutes Pre wash for 3 minutes wash for another 5 minutes and rinse it for ...

Page 20: ...cording to EN 285 to comply with EN ISO 17665 The highest sterilization temperature is 138 C The sterilization time is at least 4 minutes at a temperature of 132 C 134 C and a pressure of 2 0 bar 2 3 bars Allow a maximum sterilization time of 20 minutes at 134 C Verification of the fundamental suitability of the products for effective steam sterilization was provided by a verified testing laborato...

Page 21: ...pigments 7 2 5 Calibration is recommended when using a new other contra angle or after an extend period of operation as the running properties can change with usage cleaning and sterilization 7 2 6 Replace the battery if it seems to be running out of power sooner than it should 7 3 Transportation 7 3 1 Excessive impact and shake should be prevented in transportation Lay it carefully and lightly an...

Page 22: ...up 1 The model BAE380R uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR11 Class B The model BAE380R is suitable for used in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies b...

Page 23: ...TE I At 80 MHz end 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broad...

Page 24: ...t into root canal Keep pushing endo file near to apical foramen screen display will become normal Metal prosthesis Flush fluid touches the metal prosthesis of dental crown Endo file touches metal prosthesis Reduce flush fluid avoid contacting metal prosthesis Avoid endo file touching metal prosthesis Dental crown problem Blood oozes out to dental crown Dental crown is broken or gum hyperplasia Com...

Page 25: ...46 47 Notes Notes ...

Page 26: ...thampton NN2 6NE UK Tel 44 1604 777700 Web www bainternational com Guilin Woodpecker Medical Instrument Co Ltd Information Industrial Park Guilin National High Tech Zone Guilin Guangxi 541004 P R China MedNet EC Rep GmbH Borkstrasse 10 48163 Muenster Germany Made in China ZMN SM 433 V1 0 20211220 ...

Reviews: