30
31
Do not use chloride detergent materials.
6.2.3 Do not use bleach or chloride disinfectant materials.
6.2.4 For your own safety, please wear personal protective equipment
(gloves, glasses, mask).
6.2.5 The user is responsible for the sterility of the product for the
first cycle and each further usage as well as for the usage of damaged or
dirty instruments where applicable after sterility.
6.2.6 The water quality has to be convenient to the local regulations
especially for the last rinsing step or when using a washer-disinfector.
6.2.7 To sterilize the endodontic files, refer to the manufacturer's
instructions for use.
6.2.8 The contra-angle needs to be lubricated after cleaning and
disinfection, but before sterilization.
6.3 Cleaning and disinfection steps for the motor handpiece, the
AC adapter and the base.
Before and after each use, all the objects that were in contact with
infectious agents should be cleaned using towels impregnated with
a disinfecting and detergent solution (a bactericidal, fungicidal and
aldehyde free solution) approved by VAH/DGHM-listing, CE marking,
FDA and Health Canada.
Warning: Do not sterilize the motor handpiece, the AC adapter
and the base.
6.3.1 Pre-Op processing
Before each use, the handpiece, charger, and base must be cleaned
and disinfected. The specific steps are as follows:
Warning: The handpiece, charger, and base cannot be
cleaned and disinfected with automatic equipment. Manual cleaning and
disinfection is required.
6.3.1.1 Manual cleaning steps:
1. Take out the handpiece, charger, and base on the workbench.
2. Wet the soft cloth completely with distilled water or deionized
water, and then wipe all the surfaces of the components such as the
handpiece, charger, base, etc. until the surface of the component is not
stained.
3. Wipe the surface of the component with a dry soft nap-free cloth.
4. Repeat the above steps at least 3 times.
Note:
a)Use distilled water or deionized water for cleaning at room
temperature.
6.3.1.2 Manual disinfection steps:
1. Soak the dry soft cloth with 75% alcohol.
2. Wipe all surfaces of headpiece, charger, base and other components
with a wet soft cloth for at least 3 minutes.
3. Wipe the surface of the component with a dry soft nap-free cloth.
Note:
a) The cleaning and disinfection must be performed within 10min
before use.
b) The disinfectant used must be used immediately, no foaming is
allowed.
c) In addition to 75% alcohol, you can use non-residue disinfectants
such as Oxytech from Germany, but you must respect the concentration,
temperature and time specified by the disinfectant manufacturer.
d) After cleaning and disinfecting the handpiece, you must install a
disposable isolation sleeve before use.
6.3.2 Post-Op processing
After each use, clean and disinfect the handpiece, charger, and base
within 30 minutes. The specific steps are as follows:
Tools: Nap-free soft cloth, tray
1. Remove the contra-angle from the handpiece, place it in a clean
tray, and then remove the disposable isolation sleeve from the handpiece.
2. Soak the nap-free soft cloth with distilled water or deionized water,
and then wipe all the surfaces of the components such as the handpiece,
charger, base, etc. until the surface of the component is not stained.
3. Wet the dry soft cloth with 75% alcohol, and then wipe all surfaces
of the handpiece, charger, base and other components for 3 minutes.
4. Put the handpiece, charger, base and other components back into
the clean storage area.
Note:
a) The cleaning and disinfection must be performed within 10min
before use.
b) The disinfectant used must be used immediately, no foaming is
allowed.
c) In addition to 75% alcohol, you can use non-residue disinfectants
such as Oxytech from Germany, but you must respect the concentration,
temperature and time specified by the disinfectant manufacturer.