c) Press the Main button again, and then the motor handpiece returns
to Standby interface.
d) Hold down the Setting button “P”, then press Main button to turn
off motor handpiece. In Standby Interface, the motor handpiece would
automatically shut down after 3 minutes without any button-pressing
operation. The motor handpiece will also automatically shut down when
put on charge.
4.2 Selecting customized program sequence number
The motor handpiece has 10 memory programs(M0-M9) and 5 preset
programs, press Adjusting button “+”/“-” to change customized program
sequence number during standby state.
M0-M9 is a memory program for canal shaping and measurement,
every memory program has its own parameters such as Operation mode,
speed and torque, all these parameters can be changed.
4.3 Parameter setting
Before starting of motor handpiece, please
check the operation mode is correct.
All the parameters must be set according to files,
make sure all the parameters are correct before
starting of motor handpiece, otherwise there is
risk of file damage.
It has 5 operation modes for canal shaping
and measurement: CW, CCW, SGP, ATR and
EAL(See chapter 3.3 Terms and definition to get
the explanations of these modes.)
Press Setting button “P” once during standby
state, press Adjusting button “+”/“-” to select
correct Operation mode.
CCW mode is used to inject calcium hydroxide
and other medicant. When this mode is being
used, a double-beep sounds continuously,
used for indicating counter clockwise rotation
happening.
Repeatedly press Setting button “P” to check that all the next level
parameters of this operation mode are correct, press Adjusting button
“+”/“-” to make changes if not.
The speed setting can be adjusted from 100 rpm
to 2500 rpm.
Press Adjusting button “+”/“-” to increase or
decrease speed. Long press to fast increase or
fast decrease speed.
In ATR mode, speeds of 100~500rpm are
available.
In SGP mode, speeds of 100~500rpm are
available.
The torque setting can be adjusted from 0.4Ncm
to 5.0Ncm.
Press Adjusting button “+”/“-” to increase or
decrease torque. Long press to fast increase or
fast decrease torque.
In ATR mode, Trigger Torques of 0.4Ncm~
4.0Ncm are available.
In SGP mode, torques of 2.0Ncm~5.0Ncm are
available.
Actions that happen automatically when the file
tip reaches the point inside the canal determined
by the Flash Bar setting.
The benefit of integration of length
determination is that when the file reaches
the reference point, the motor will response
according to setting. It can be Reverse , Stop
and OFF.
P ress Adjusting button “+”/“-” to change.
OFF: Disable Apical Action function, file rotates
as usual even if it reaches the reference point.
Stop: automatic rotation stop when reference
point is reached, pull upward a little bit and file
will rotate again.
Reverse: automatically reverses rotation when
the file reaches or passes the reference point;
pull upward a little bit, the rotation direction
will change back again.