Chapter 6.
MAINTENANCE AND TROUBLESHOOTING
33
■
Maintenance
Be sure to confirm that the
MCF
is operating normally by inspecting it periodically
once or more per year. The
MCF
has a self-diagnosis function. If an alarm code
does not disappear after taking countermeasures, contact the azbil Group or your
dealer. Repair or replacement might be necessary.
●
Measurement module parts
•
Side view of measurement module
•
View without measurement module
●
Maintaining the measurement module
Required items: Allen wrench (M5), Air blower, protective goggles, helmet,
gloves.
Handling Precautions
• For your safety, wear goggles while taking the measurement module
off.
Bottom view
(View without rubber sheet)
Bottom view
Rubber sheet
Flow path of
measurement module
Built-in filter
Flow direction arrow
Case retaining screws
(Do not remove)
CAUTION
When carrying the flowmeter or connecting it to the pipe, do not hold it by the measurement
module. Doing so could cause damage, or the device could drop, causing an injury.
When maintaining or replacing the measurement module, release the internal pressure of this
device before removing the measurement module.
Bypass inlet / outlet
Top view of the main path
Side view of the main path
Packing
Flow direction arrow
Flow path identifier (5 digits 3)
XXXXX-XXXXX-XXXXX