Chapter 5. SETTING AND OPERATION
■
Device information display
●
Viewing the display
(1) Press and hold the
[enter]
key for 2 seconds or more while the instantaneous
flow rate or integrated flow is indicated on the display.
>> The
MCF
moves to the device information display mode, and indicates the
item and the related information alternately.
(2) To move to the next item, push
[
]
. To return to the previous item, push
[
]
.
(3) To quit the device information display mode, push the
[mode]
key.
>> The
MCF
returns to normal indication mode.
●
Device information
*1: The conversion factor can be changed by changing the cost rate (
COSt
) setting in the parameter setup.
In the case of integrated flow countdown, the cost is not indicated.
.
Item
Name
Description
bAtt
Battery level
Indicates the battery level.
"
O O O
": Full charge
"
O O
": Good
"
O
": Low
"
-
": Critical
Id_0 1
Model number ID
Used for model identification.
MCF0080
→
0
MCF0150
→
1
MCF0151
→
5
MCF0250
→
2
MCF0400
→
3
MCF0500
→
4
Id_02
Range ID
Used to identify standard ranges
MCF0080
→
2000 (200.0L/min)
MCF0150
→
5000 (500.0L/min)
MCF0151
→
10000 (1000.0L/min)
MCF0250
→
3000 (3000L/min)
MCF0400
→
6000 (6000L/min)
MCF0500
→
12000 (12000L/min)
Id_03
Firmware version
COSt.L
Cost (last digits)
Displays the cost, calculated from the integrated flow
COSt.H
Cost (first digits)
value. *
1
ItOt.L
Total integrated flow (last digits)
Displays the integrated flow value since the time of
ItOt.H
Total integrated flow (first digits)
shipping. Even if the integrated value is reset, this
value will not return to 0.
IPrE.L
Integrated flow before resetting (last digits)
The integrated flow before it was reset.
IPrE.H
Integrated flow before resetting (first digits)