Balance checking
It is advised to check scale indication accuracy before and after series of
measurement using any load with known weight.
To check the scale with legal verification use a calibration weight with valid
calibration certificate. In case permissible error is exceeded it is advised to contact
the nearest service to calibrate the scale.
11.
Scale adjustment
Scale adjustment (calibration with external standard of mass) should be made, if
the scale precision is unsatisfactory. In that situation M1 standard of mass should
be used (according to OIML) with the value equal to scale Max range (technical
data) and a special function
CALIb
detailed in 14.18 chapter.
CALIb
function is only available in scales not designed for legalization.
To adjust a balance it is necessary to break protective marks, therefore
to calibrate the scale please contact the nearest service.