42
USER MANUAL
_____________________________________________________________________________________________________________________
14.20 Entering reference zero function (Zero)
Note:
This function is enabled in non legalized scales only.
ZER function allows entering new value of reference zero (value referred to empty
pan) without need of contacting with authorised service centre.
Press
MENU
key.
When
Zero
is displayed press
→
T
←
key.
The
following
options
appear
successively on display:
ZEr-0
– activate function,
ZEr-1
– enter new zero value,
ZEr-2
– enter new protecting code.
Using
→
T
←
key, choose
ZEr-1
and
enter access code for function (in new
scale it is 1234). Direct result from A/C
converter will appear on scale display.
When the pan is empty press
→
0
←
key.
Wait for finishing zeroing process.
In order to change access code use
ZEr-2
option. Entering value is
performed similarly as with
ZEr-1
option.
Err-b
ZEr-1
ZEro
T
T
-
1234
1
12
12
5007
0.00
T
T
...
...