14.2  Autozeroing function  (AutoZE)  

 

When 

F..-Aut

  function  is  activated, 

the scale automatically ensures stable 
zero indication if the pan is empty or 
if  zero  indication  was  acquired  by 
pressing 

T

 key. 

To  leave  the  function  press 

MENU

 

key,    then    with   

T

    key  chose 

AutoZE

 and 

Aut-0

 
Note:  

Autozeroing  function  is  activated 
automatically  for  10  min.  after 
switching-on. 

 

 

 

 

0.01

AutoZE

Aut-1

Aut-0

0.00

T

T

Summary of Contents for B/P Series

Page 1: ...USER MANUAL B P SERIES FILE 2010 10 BP 113 B 50_0 GB...

Page 2: ...ighing 16 13 2 Weighing with tare 17 13 3 Increased readability 17 14 Special function describes 18 14 1 Menu customization function ACtIV and dEFAUL 19 14 2 Autozeroing function AutoZE 20 14 3 Pieces...

Page 3: ...umber LAbEL 39 14 18 External calibration CALIb 40 14 19 Setting time of stabilization function Stb 41 14 20 Entering reference zero function Zero 42 14 21 Automatic switching off the scale function A...

Page 4: ...rotecting scale casing against opening notified body stamps and green metrological marking placed on the balance name plate Renewing of scale legal verification is required when protective seal is vio...

Page 5: ...a Scale type B30P B50P Max load 30kg 50kg Min load 200g 400kg Readout graduation d 10g 20g Verification graduation e 10g 20g Tare range 30kg 50kg Accuracy class III Working temperature 10 40 C Weighin...

Page 6: ...uk wyniku klawisz HR zwi kszenie rozdzielczo ci wskaza masy wska nik 0 wska nik wyzerowania przy nieobci onej wadze wska nik sygnalizuje ustabilizowanie si wyniku wa enia wska nik NET masa netto po u...

Page 7: ...e type supplied with the scale Pay attention that supply voltage is compatible with specified technical data Do not use the scale when its cover is opened Do not use the scale in explosive conditions...

Page 8: ...ements from below the pan 3 Place the scale on a stable ground not affected by mechanical vibrations and airflows 4 Level the scale with the rotating rear legs so that the air bubble in the water leve...

Page 9: ...30V 50Hz and plug the feeder connector to the 12V power socket at the back of the scale The scale proceeds with following start up actions Display test Basic electronic components tests C1 C 2 C 6 Pro...

Page 10: ...d with a load indicator graduated in percentages 3 Weighing result should be read when the indicator lights which signalises stabilisation of a result 4 When the scale is not used but it is necessary...

Page 11: ...r is exceeded it is advised to contact the nearest service to calibrate the scale 11 Scale adjustment Scale adjustment calibration with external standard of mass should be made if the scale precision...

Page 12: ...is chose function LPt a printer data is send automatically after result stabilisation but next transmission is possible after removing previously weighted sample The scale sends following information...

Page 13: ...13 g b t c or Byte 14 space Byte 15 CR Byte 16 LF Attention Network number different than zero F rS nr function changes scale working mode communication with a computer is possible after logging the...

Page 14: ...e digits dot decimal separator number of digits after dot should be the same as on scale display Scale Computer without response Example in order to set low threshold 1000g in scale B1 5 d 0 5g the fo...

Page 15: ...ls starting instruction hh mm 5 signs hours minutes yyyy mm dd 10 signs year month day mass unit 10 signs scale indicate mass unit P1 ending instruction Attention 1 It is possible to place a constants...

Page 16: ...on in further part of instruction descriptions are replaced with pictures put a load on the pan remove the load from the pan press the key when indication is displayed forced change automatic change 1...

Page 17: ...display gross weight press B G key Note Press B G key to return to net weight indication 13 3 Increased readability Press HR key to display the weighing result for 5s with the highest readability poss...

Page 18: ...function AVErA percentage weighing function PErC extended calibration function CALIb setting time of stabilisation function Stb selecting label number function LabEL automatic switching off scale fun...

Page 19: ...ration time longer Operation sequence shown in the pictures on the left causes adding function for setting serial interface RS232C parameters Port to function menu After switching on ACTIV function a...

Page 20: ...AutoZE When F Aut function is activated the scale automatically ensures stable zero indication if the pan is empty or if zero indication was acquired by pressing T key To leave the function press MENU...

Page 21: ...unit and sample weight used in first phase is bigger than 100 reading units To leave function press MENU key and then with key chose PCS and PCS 0 Note 1 Err 3 communicate signalises that a sample wa...

Page 22: ...ial port working mode cooperation with a printer After activation the scale prints a header Weighing result with a successive measurement number is printed automatically after result stabilisation wit...

Page 23: ...ontrol PArItY 0 1 Odd 0 1 network number when working in multistand computer system when working as a single scale the value should be 0 continuous transmission without using key approx 10 results per...

Page 24: ...ntered using the keypad or by sampling container weight from the pan Operation sequence The following options are possible tAr 0 leave the function tAr 1 activate the function with the previous tare v...

Page 25: ...hing with constant tare When tAr function is activated press 0 key to zero the indication and to recall tare value from the memory Tare value is displayed with sign B G key enables for instant switchi...

Page 26: ...ing results into summing register for average calculation Measurement registration is performed automatically in the moment when scale indication becomes stable Short time displaying of denotes that l...

Page 27: ...uccessive ingredients A B C etc are weighed each time starting from zero indication In order to allow this after weighing of each ingredient tare the scale After weighing of several ingredients readin...

Page 28: ...is displayed in different format depending on the weight value of reference sample For weight values of reference sample 0 3 5 of weighing range result is displayed in format 100 for range 3 5 35 in f...

Page 29: ...ical vibrations air blasts base vibrations on measurement result Press MENU key and select FILtEr pressing T key The following options will be shown successively on display FIL 0 work without FIL 1 fi...

Page 30: ...mum value shown by the scale in a while Before measurement scale should be tared After using MENU key and selecting UP function the highest mass result will be hold on display Pressing T key will caus...

Page 31: ...____________________________________________________________ 14 12 Force measuring function nEWto Function activation causes displaying result in force units N Press MENU key Using T key choose NEWto...

Page 32: ...1 is displayed press T key Tare the scale using T key if necessary and place the animal on the pan Wait until the weighing result is averaged scale display will be blinking Then scale will show stable...

Page 33: ...r threshold the scale signals MAX red colour long sound signal The checkweighing results can be use to control optical indicator Indication mode batching devices Batching mode Standard scale is set fo...

Page 34: ...odes e g 1N4148 It is advised to use MS3K P electronic board sold separately consisting of RM96P transmitters with DC24V input voltage and AC250V 3A output Important notes 1 After switching the scale...

Page 35: ...esholds are displayed SEt 0 go to weighing with signalling threshold excess SEt 1 set lower threshold value SEt 2 set upper threshold value SEt 3 set zero signalisation threshold Using T key select SE...

Page 36: ...printout and leaving the function tot 1 working with receipt printout after each measurement tot 2 working without receipt printout Press T key when tot 1 is displayed Perform measurement series pres...

Page 37: ...rnal clock Date and Time do not appear on printout Maximum number of measurements 99 999 Maximum total load 99 999 000d The weighing unit of the total value from the register Total is the same as the...

Page 38: ...e during printout of current indication key dAt 2 change current date and time The example at the left presents how to set current date and time using dAt 2 option On successive positions digits are c...

Page 39: ...nter memory according to printer user manual Choosing label form is performed by entering label number using LAbEL function Press MENU key When LAbEL is displayed press T key The following options app...

Page 40: ...peration sequence Press MENU key to call functions menu and to choose CALIbr with T key The following options appear successively on display CAL 0 leave without calibration CAL 1 quick calibration wit...

Page 41: ...intout on a printer connected to the scale Press MENU key When Stb is displayed press T key The following options appear successively on display Stb 0 deactivate the function Stb 1 the longest stabili...

Page 42: ...vice centre Press MENU key When Zero is displayed press T key The following options appear successively on display ZEr 0 activate function ZEr 1 enter new zero value ZEr 2 enter new protecting code Us...

Page 43: ...tare values are remembered in scale memory After next start up of the scale these values are restored Press MENU key When AutoOF is displayed press T key The following options appear successively on...

Page 44: ...ther than distilled water choose dEnSIt from menu and use dEn 2 option to enter liquid density taking into consideration its temperature To enter value use the following keys 0 digit increase decimal...

Page 45: ...of solid density measurement results will be performed in the following form It is recommended to use pan hanging below balance weighing in the air and in liquid is performed then in the following wa...

Page 46: ...ut on and scale indication is stable Each time after load is put on printout is performed measurement number result date and hour Accepting next measurement is possible after taking last load off For...

Page 47: ...n F 3 is displayed The following options appear successively on display O_n nominal mass Qn 20 60 100 160 n select available measurement quantity total quantity of samples out exit 8 Press T key when...

Page 48: ...rinted 19 In order to finish working with the function and reset results register press MENU key and when F tP and F 0 is displayed press T key Note Activating TP function causes that indicators signa...

Page 49: ...e evaluates n number of samples sum x sum of all samples n x x sum _ x average value sum x n min minimal value from n samples max maximal value from n samples max min maximal value minus minima value...

Page 50: ...er allowable lower value NG amount of measurements above allowable upper value To finish work with this function and zeroing result register press F key then during F StA and F 0 is displayed press T...

Page 51: ...0 20 50 100 Press MENU key to access Function Menu To enter the function press T key when F PAP is displayed n is the number of samples placed on the pan To choose previously used value select option...

Page 52: ...ll repairs must be performed by the authorized personnel 6 To perform the repair contact the nearest service point Emergency messages Message Reason Recommendation C 1 6 over 1min autotest negative re...

Page 53: ...directive electromagnetic compatibility Moreover balances with the following information on rating plates sticker with two digits of year of conformity assessment and with number of notified unit perf...

Reviews: