INSTRUCTION MANUAL
15
_____________________________________________________________________________________________________________________
Scale zeroing (calling
0
key press):
Computer
Scale:
SZ
CR LF (53h 5Ah 0Dh 0Ah),
Scale
Computer: without response,
Scale turning on / off (calling I/ key press):
Computer
Scale:
SS
CR LF (53h 53h 0Dh 0Ah),
Scale
Computer: without response,
Entering to special function menu (calling
MENU
key press):
Computer
Scale:
SF
CR LF (53h 46h 0Dh 0Ah),
Scale
Computer: without response,
Setting threshold 1 value (option):
Computer
Scale:
SL
D1...DN
CR LF (53h 4Ch
D1...DN
0Dh 0Ah)
D1...DN
– threshold value, maximum 8 characters („-” – negative value, digits, dot – decimal
separator), number of digits after dot should be the same as on scale display,
Scale
Computer: without response,
Example:
in order to set low threshold 1000g in scale B1.5 (d=0.5g) the following order should be sent:
S L 1 0 0 0 . 0 CR LF (53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah),
in order to set low threshold 100kg in scale B150 (d=50g) the following order should be sent:
S L 1 0 0 . 0 0 CR LF (53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),),
Setting threshold 2 value (option):
Computer
Scale:
SH
D1...DN
CR LF (53h 48h
D1...DN
0Dh 0Ah),
D1...DN
– threshold value, maximum 8 characters
Scale
Computer: without response.
Setting threshold 3 value (option):
Komputer
Waga:
SM
D1...DN
CR LF (53h 4Dh
D1...DN
0Dh 0Ah),
gdzie:
D1...DN
– threshold value, maximum 8 characters
Waga
Komputer: without response.
Connecting cable WK-1
(scale – computer / 9-pin interface):
Connection cable WD-1
(connects scale with AXIS printer):
SCALE
KAFKA
PRINTER
SCALE
COMPUTER