40
INSTRUCTION MANUAL
_____________________________________________________________________________________________________________________
14.20 Entering reference zero function (Zero)
Note:
This function is enabled in non legalized scales only.
ZER function allows entering new value of reference zero (value referred to empty
pan) without need of contacting with authorised service centre.
Press
MENU
key while
Err-b
is
displayed.
Zero
function
will
be
activated
automatically.
After
CodE
communicate
disappears
insert access code to function (in new
scale 1234).
The following options
appear successively on display:
ZEr-0
– activate function,
ZEr-1
– enter new zero value,
ZEr-2
– enter new protecting code.
Using
→
T
←
key, choose
ZEr-1
and
direct result from A/C converter will
appear on scale display.
When the pan is empty press
→
0
←
key.
Wait for finishing zeroing process.
In order to change access code use
ZEr-2
option.
Entering
value
is
performed similarly as with
ZEr-1
option.
It is also possible to use function in
standard way:
Activate
Zero
function using
ACtIV
function.
Press
MENU
key.
When
ZEro
is displayed press
→
T
←
key and so on.
Err-b
CodE
ZEr-1
T
-
1234
1
12
12
40007
- -
0.00
T
T
...
...