![Avtech AVR-G1-B-P-EMRA-R5-RP-IM Instructions Manual Download Page 23](http://html1.mh-extra.com/html/avtech/avr-g1-b-p-emra-r5-rp-im/avr-g1-b-p-emra-r5-rp-im_instructions-manual_3034550023.webp)
OPERATIONAL CHECK
This section describes a sequence to confirm the basic operation of the instrument. It
should be performed after receiving the instrument. It is a useful learning exercise as
well.
Before proceeding with this procedure, finish read this instruction manual thoroughly.
Then read the “Local Control” section of the “Programming Manual for -B Instruments”
thoroughly. The “Local Control” section describes the front panel controls used in this
operational check - in particular, the MOVE, CHANGE, and ADJUST controls.
1. Connect a cable from the SYNC OUT connector to the TRIG input of an
oscilloscope. Connect a 11W (or higher) 1.6 kΩ load to the OUT connector and
place the scope probe across this load. The load resistor must have a voltage
rating of at least 240V. The power dissipated in the resistor is given by
P = (V
2
/ R) × (PW / T) = (V
2
/ R) × PW × f
where “V” is the output voltage, “R” is the load resistance, “PW” is the pulse width,
and “T” is the pulse period (1/frequency), and “f” is the frequency.
2. Set the oscilloscope to trigger externally with the vertical setting at 50 Volts/div and
the horizontal setting at 1 us/div. Be sure that your oscilloscope and probe setup
can handle the maximum amplitude of 240V. The load resistor should be rated for
at least 240V of voltage and 20W of power.
3. Turn on the AVR-G1-B-P-EMRA-R5-RP-IM. The main menu will appear on the
LCD.
23
AVTECH
PULSER
MAIN OUTPUT
CONNECTOR
SYNC
OUTPUT
AC
POWER
ALL CABLES: 50 OHM COAXIAL
REAL-TIME
OSCILLOSCOPE
CHANNEL A
TRIG
INPUT
SCOPE
PROBE
TEST
LOAD
Summary of Contents for AVR-G1-B-P-EMRA-R5-RP-IM
Page 32: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...
Page 39: ...PERFORMANCE CHECK SHEET 39...