Routine Maintenance
17.
When
appears, the system pressurizes and flushes for 400 seconds (6 minutes and
40 seconds).
18.
When
is displayed, touch
to return to the main menu.
19.
Replace the left front panel.
20.
Dispose of the used purification pack following your site requirements for disposal.
Replace Air Filter on External Storage Tank
All tanks use air filter (
TC005
). The air filter connects to the top of the storage tank.
Follow this procedure.
1.
Remove the new air filter from its packaging.
2.
Pull the existing filter off the top of the tank.
3.
Turn the new filter so the flange-end faces down toward the filter-connection tube as shown
below.
4.
Place the filter over the tube on the tank. and press down.