Geno Range RO/DI Water System, Series 5 Operation Manual
Avidity Science, LLC
10
Operation
Warning Messages
Below are the warning messages and solutions.
Description
Error
Number
Solution
Water leak (system shuts down)
60
Fix the leak. Dry the leak sensor located on the bottom of
the cabinet.
Contact Technical Support as necessary.
Permeate
a
a. When RO pump is ON
40
Contact Technical Support.
DI Quality Low
b
b. When Permeate is ON
41
Follow on-screen instructions.
Loop Resistivity
c
c. When circulation pump is active
42
Follow on-screen instructions.
Loop Temperature
43
Follow on-screen instructions.
Tank Empty
50
Follow on-screen instructions.
Tank Low
51
Tank level is 30 percent or less.
If tank water level displays
-71%
, check all connections to
the storage tank and the machine.
DI Quality Sensor Fault: LOW
62
Follow on-screen instructions.
Loop Conductivity Sensor Fault
63
Follow on-screen instructions.
Temperature Sensor: HIGH
64
Allow system to flush.
Temperature Sensor: LOW
65
Check incoming water temperature.
Level Sensor Fault:
HIGH
67
Contact Technical Support.
Level Sensor Fault: LOW
68
Contact Technical Support.
Tank UV Lamp Run Time Expired
5
Contact Technical Support
Pretreatment Module TC001 time expired
1
Replace.
Endure Purification Pack 1 TC002 time
expired
2
Replace.
Inlet Conductivity: HIGH
24
Check conductivity in Feed Water.
Pressure Sensor Fault: HIGH
69
100-litre storage tank with pump pressure sensor
Contact Technical Support
Pressure Sensor Fault: LOW
70
100-litre storage tank with pump pressure sensor.
Contact Technical Support
Pressure: HIGH
71
Geno pressure sensor. Follow instructions on the screen.
Contact Technical Support if necessary.