For proper weight reading, the signals reaching the indicator must be
identical no matter where a weight is placed on the scale. Getting these
signals to match is called corner balancing the scale.

Your goal is to get the readings from the weight sensors to match. You do
not have to get the correct weight reading at this point. That is taken care of
when you calibrate your indicator. (See the 

Service Manual for your indica-

tor.) One potentiometer affects one weight sensor. You balance the weight
sensors by adjusting the corresponding potentiometer in the junction box.

This scale was corner balanced at the factory, but in a new installation it is
required that corner balancing and calibration be checked to ensure installa-
tion accuracy. Below are the steps you need to take to corner balance your
scale:

1a. Test load the deck in both directions by driving a loaded truck on and off

the scale 3 or 4 times.

1b. Check all mechanical hardware.

1c. Remove the junction box cover to access the potentiometers.

2.

To capture the value of internal zero for your particular indicator, refer to
your indicator's 

Service Manual.

3.

Use test weights equal to 20-25% of full capacity and obtain a displayed
weight value for the test weight applied to each of the four weight
sensors, like this:

3a. Disable AZT on indicator.

3b. Place certified test weight directly above first weight sensor.

3c. Record displayed weight value.

3d. Repeat steps 3b and 3c for each weight sensor.

4.

If displayed weight values for all weight sensors equal each other, within
+/- 1 division, proceed to the calibration instructions for your indicator
found in

5.

If displayed weight value for any weight sensor varies from the others by
more than +/- 1 division, adjust the appropriate junction box potentiom-
eter by turning it the number of 360 degree turns indicated by this
formula:

If the Number Of Turns is a positive value, turn the potentiometer
clockwise. If Number Of Turns is a negative value, turn the potentiom-
eter counterclockwise.

6.

Repeat steps 3b and 3c followed by step 4 or step 5 until all weight
sensors equal each other, /- 1 division.

Corner Balancing

Summary of Contents for BridgePort

Page 1: ...profile self contained portable scale suited for a wide variety of applications It is designed and manufactured for simplified temporary installation and relocation when necessary The basic units are...

Page 2: ...6 below the frost line To speed the installation of the scale ensure the proper installation of the piers before receiving the scale For the dimension information refer to the installation drawing Th...

Page 3: ...ock on the frame Adjust the check bolts to a maximum of 3 16 clearance Refer to Figure 3 Figure 3 Checkbolts Install the optional bulkheads on each end of the scale If the bulkheads are from Weigh Tro...

Page 4: ...d level for a minimum of 10 feet at each end of the scale The ground rods must be installed as close as possible to the scale module at each end and same side as the junction box Connect the ground ro...

Page 5: ...ll the T gap covers optional at each end of the scale to cover the gap between the edge of the scale deck and the bulkhead The scale is now ready for testing calibration and inspection as required Fig...

Page 6: ...n Position each module end to end Make sure the junction boxes remain on the same side of the scale To ensure the proper positioning of the second and third module if appli cable the units must be as...

Page 7: ...ulkhead Weigh Tronix will not be responsible for error caused by other bulkheads Once the bulkheads are properly installed prepare the ramps at each end of the scale The ramps must be straight and lev...

Page 8: ...box connectors Remove the tie down bolts and store them in a safe place for future use Install the hold down bolts to secure the scale weigh bridge to the frame To prevent movement of the scale platf...

Page 9: ...gh bar 15 std cable 271 73 0046 Weigh bar 50 std cable 271 73 0053 Weigh bar 15 S S sheathed cable 27173 0061 Weigh bar 50 S S sheathed cable 27173 0079 50 home run std cable 1 7 pin male connector 47...

Page 10: ...ction box cover to access the potentiometers 2 To capture the value of internal zero for your particular indicator refer to your indicator s Service Manual 3 Use test weights equal to 20 25 of full ca...

Page 11: ...A Weigh Tronix 1000 Armstrong Dr Fairmont MN 56031 USA Telephone 507 238 4461 Facsimile 507 238 4195 e mail industrial weigh tronix com www weigh tronix com Weigh Tronix Canada ULC 217 Brunswick Blvd...

Reviews: