background image

Instruction Manual QNix

®

8500 

- 16 - 

of the display, the wireless symbol is indicated together with a small “P” for probe. You can 
take measurements with the radio controlled probe the same way you do with the usual 
probes. Each reading that is correctly transmitted to the gauge is indicated with a short flash-
ing of the blue probe light. That way you can observe if the probe is still inside radio range.  

 

As the radio-controlled probe uses a special radio mode to ensure maximum operating safe-
ty, observe the following during radio communication between gauge and probe:  

With disconnected probe, the wireless mode is switched to the probe and any communica-
tion between gauge and PC is no longer possible. However, readings are still transmitted to 
the PC as online-measurements. To readout the memory and to change the gauges setting 
via PC the probe has to be reconnected to the gauge.  

Deactivating the wireless mode via the gauges menu has no effect on the radio-controlled 
probe. An activation of the wireless mode using the gauge however leads to a disconnection 
of the communication with the radio-controlled probe.  

10 

Troubleshooting 

10.1 

Incorrect or dubious readings 

If the gauge shows incorrect or dubious readings, please check: 

  Does the display indicate the correct substrate? If you are not sure whether your sub-

strate is Fe or NFe, test it with a magnet; if the substrate is attracted by the magnet, 
choose the an Fe probe or the Fe measuring mode of a dual probe, otherwise choose 
the NFe mode or a NFe probe. 

  Is the Zero Reference correctly set? You achieve the best results when you zero the 

probe on an uncoated piece of the substrate whose coating you want to measure. If 
an uncoated piece is unobtainable you can use the standard plates that are supplied 
with your gauge. Refer to §6 for further details. Please note that it is the probe that 
you zero, not the gauge. The results of the zeroing are stored in the probe. Thus if 
you switch the probe, you might have to repeat the setting of the zero reference. 

  Is the substrate you want to measure suitable for a standard calibration? Certain al-

loys that contain iron, in particular some kinds of stainless steel, require special con-
sideration. You should use a custom calibration ( see §8.1.1) in these cases. 

10.2 

Automatic selection of the substrate does not work 

This is caused by an incorrect setting of the zero references of the probe. To recover from 
this, follow these steps: 

1)  Set the gauge to Fe mode (select Fe from the substrate menu, see §7). 

2)  Set the zero reference on an uncoated sample of Fe material or on the standard Fe 

plate that was supplied with your gauge. 

3)  Set the gauge to NFe mode (select NFe from the substrate menu). 

4)  Set the zero reference on an uncoated sample of NFe material or on the standard 

NFe plate that was supplied with your gauge. 

10.3 

My gauge does not provide all the described features 

There are 2 possible reasons for that:  

You have the Basic version and the feature you are looking for is only available in the Premi-
um version. Check the menu overview on page 2 and 3. 

The feature you are looking for was introduced after you purchased your gauge. In that case 
an update of the gauge firmware might help. Please contact your dealer for details. 
If you want to know the version of the firmware in your gauge, select “Info” from the setup 
menu (see §8.5.3). The gauge will display a popup window containing the serial number of 
the gauge and a 7 digit string indication hardware and software version of the gauge. The 

Summary of Contents for QNix 8500

Page 1: ......

Page 2: ...Instruction Manual QNix 8500 2...

Page 3: ...Instruction Manual QNix 8500 3...

Page 4: ...Averaging 11 8 3 Data Memory 11 8 3 1 Delete All 12 8 3 2 Delete values 12 8 3 3 Data Memory Mode ON OFF 12 8 3 4 Creating a New Batch 12 8 3 5 Selecting Editing a Batch 12 8 4 Statistics 13 8 4 1 Se...

Page 5: ...e PC Software 17 14 Loading a new Language into the Gauge 17 15 Updating the Software in the Gauge and Probe 18 16 Display Symbols 18 17 Scope of Delivery 19 18 Technical Data 20 Notice This instructi...

Page 6: ...to strong magnetic fields Remove the batteries from the gauge and store them separately if the device is not used for a longer period The gauge is not waterproof Do not expose it to water 3 Changing...

Page 7: ...ring head Within the measuring range of the probe measurements like e g distance measurements are possible that way 6 Setting the Zero Reference When using the gauge for the first time after changing...

Page 8: ...u For your convenience find a diagram of the menu structure showing all possible menu op tions on the first two pages of this instruction manual Depending on the version of your gauge only a part of t...

Page 9: ...d calibrations under the menu item Define New Calibration To select a certain calibration use the and buttons Subse quently pressing opens the corresponding options menu A dot at the right side of the...

Page 10: ...a number of zero reference readings max 19 to be taken with the probe before the gauge will calculate the average zero reference based on the readings taken After selecting the menu item pressing the...

Page 11: ...t To set or delete the lower limit please follow the instructions given under 8 2 3 Readings below the lower limit are indicated in the display with the symbol and with an au dio warning signal 8 2 5...

Page 12: ...firm it by pressing For corrections of prior digits simply press the backspace button at any time NOTE When creating a new batch after other batches i e B 0001 B 0002 have been de leted the gauge stil...

Page 13: ...he statistics functions are calculated per memory block The gauge determines the following statistical parameters average value standard deviation maximum minimum Within the statistics menu two differ...

Page 14: ...ing of the display To change the display orientation press the button then select the setting Normal or Flip using and and finally confirm by pressing again NOTE When the display is flipped the functi...

Page 15: ...he and buttons Confirm with For corrections use the backspace button 8 5 6 2 Setting the Date Set the date the same way you set the time 8 5 7 The button The button is used to set the zero reference s...

Page 16: ...coating you want to measure If an uncoated piece is unobtainable you can use the standard plates that are supplied with your gauge Refer to 6 for further details Please note that it is the probe that...

Page 17: ...tion guide on the CD 13 Setting the Gauge Options with the PC Software The setup options described in 6can also be programmed and then downloaded into the gauge using the PC software This is especiall...

Page 18: ...rror operating error Infi readings outside measuring range wrong type of surface readings exceeding upper limit readings below lower limit x average standard deviation wireless interface active p radi...

Page 19: ...th clip Plastic case for transportation and storage Optional Accessories Fe probe 2 mm Fe probe 5 mm NFe probe 2 mm Dual probe Fe NFe 2 2 mm Dual probe Fe NFe 5 2 mm Calibration films zero reference p...

Page 20: ...max 10 m on the free field Temperature range Storage 10 C to 60 C 14 F to 140 F Operation 0 C to 50 C 32 F to 122 Power supply 2 x batteries AA 1 5V alkaline or 2 x rechargeable batteries AA 1 2V Dime...

Page 21: ...e pole and negative pole to avoid short circuiting the device Do not use new and old batteries or batteries of a different kind together This could lead to miss functions in your device Remove empty b...

Reviews: