background image

Instruction Manual QNix

®

8500 

- 14 - 

8.4.2 

Displaying a summary of statistical data 

If you select a block that is not active, a summary of statistical data will be displayed immedi-
ately. If you select the active block, choose 

„Show Values” from the submenu to display the 

summary. 

8.5 

Gauge Setup 

8.5.1 

LC-Display 

8.5.1.1 

Unit of a Reading  

Readings can be displayed either in µm or in mil (standard setting: µm).  
Select the unit pressing the ▲ and ▼ buttons and then confirm with ►. 

8.5.1.2 

Resolution  

Setting the decimal digits adjusts the resolution of the readings to the application. It can be 
useful for example to blank out the decimal places when working on surfaces with high 
roughness leads to high variations in the readings. The resolution can be 

set to „Coarse“ 

(0 

µm), „Normal“ (0.0 µm) or „Fine” (0.00 µm).   

8.5.1.3 

Backlight 

In 

„Auto" mode, the backlight is turned on for about 7 seconds while taking a reading. Switch 

off the backlight using the menu item „OFF“. 
Switching the backlight off will increase the service life of the batteries.  

8.5.1.4 

Display Orientation 

The display orientation can be adjusted to the way the gauge is operated.   

If the probe is connected to the gauge via adapter cable and the gauge is placed on a table, 
the display orientation can be flipped by 180° to allow for an upright reading of the display.  

To change the display orientation, press the 

► button then select the setting “Normal” or 

“Flip” using ▲ and ▼ and finally confirm by pressing ►again.  
NOTE: When the display is flipped, the functions of the keypad are also flipped. As soon as 
you confirmed the flipping, you should turn the gauge by 180° in order to operate the gauge 
in a familiar manner.  

Leave the menu item by pressing the (backspace) button: 

◄.

 

8.5.2 

Wireless Interface  

If the wireless interface is activated, the gauge is constantly ready to receive commands/data 
from the PC-Software. For example the gauge can be programmed for specific applications 
using this interface (see § 12). You can also update the gauge software this way.   

When the wireless interface is active, the antenna symbol 

Ψ

 appears in the lower left hand 

corner of the display. In addition, the red LED in the keypad blinks every 7 seconds (even 
when the gauge is turned off).  

If data is transmitted via the wireless interface, the red LED flashes.  

Deactivating the interface while not in use will increase the service life of your batteries. If the 
gauge is not used for more than 30 minutes, the interface will deactivate itself automatically. 
The next time the gauge is turned on the interface is activated again.  
Using the “EnergySave” mode via the setup menu will deactivate the interface together with 
the display (after about 30 seconds). This increases the service life of your batteries com-
pared to the setting “ON” depending on the application conditions. 

Summary of Contents for QNix 8500

Page 1: ......

Page 2: ...Instruction Manual QNix 8500 2...

Page 3: ...Instruction Manual QNix 8500 3...

Page 4: ...Averaging 11 8 3 Data Memory 11 8 3 1 Delete All 12 8 3 2 Delete values 12 8 3 3 Data Memory Mode ON OFF 12 8 3 4 Creating a New Batch 12 8 3 5 Selecting Editing a Batch 12 8 4 Statistics 13 8 4 1 Se...

Page 5: ...e PC Software 17 14 Loading a new Language into the Gauge 17 15 Updating the Software in the Gauge and Probe 18 16 Display Symbols 18 17 Scope of Delivery 19 18 Technical Data 20 Notice This instructi...

Page 6: ...to strong magnetic fields Remove the batteries from the gauge and store them separately if the device is not used for a longer period The gauge is not waterproof Do not expose it to water 3 Changing...

Page 7: ...ring head Within the measuring range of the probe measurements like e g distance measurements are possible that way 6 Setting the Zero Reference When using the gauge for the first time after changing...

Page 8: ...u For your convenience find a diagram of the menu structure showing all possible menu op tions on the first two pages of this instruction manual Depending on the version of your gauge only a part of t...

Page 9: ...d calibrations under the menu item Define New Calibration To select a certain calibration use the and buttons Subse quently pressing opens the corresponding options menu A dot at the right side of the...

Page 10: ...a number of zero reference readings max 19 to be taken with the probe before the gauge will calculate the average zero reference based on the readings taken After selecting the menu item pressing the...

Page 11: ...t To set or delete the lower limit please follow the instructions given under 8 2 3 Readings below the lower limit are indicated in the display with the symbol and with an au dio warning signal 8 2 5...

Page 12: ...firm it by pressing For corrections of prior digits simply press the backspace button at any time NOTE When creating a new batch after other batches i e B 0001 B 0002 have been de leted the gauge stil...

Page 13: ...he statistics functions are calculated per memory block The gauge determines the following statistical parameters average value standard deviation maximum minimum Within the statistics menu two differ...

Page 14: ...ing of the display To change the display orientation press the button then select the setting Normal or Flip using and and finally confirm by pressing again NOTE When the display is flipped the functi...

Page 15: ...he and buttons Confirm with For corrections use the backspace button 8 5 6 2 Setting the Date Set the date the same way you set the time 8 5 7 The button The button is used to set the zero reference s...

Page 16: ...coating you want to measure If an uncoated piece is unobtainable you can use the standard plates that are supplied with your gauge Refer to 6 for further details Please note that it is the probe that...

Page 17: ...tion guide on the CD 13 Setting the Gauge Options with the PC Software The setup options described in 6can also be programmed and then downloaded into the gauge using the PC software This is especiall...

Page 18: ...rror operating error Infi readings outside measuring range wrong type of surface readings exceeding upper limit readings below lower limit x average standard deviation wireless interface active p radi...

Page 19: ...th clip Plastic case for transportation and storage Optional Accessories Fe probe 2 mm Fe probe 5 mm NFe probe 2 mm Dual probe Fe NFe 2 2 mm Dual probe Fe NFe 5 2 mm Calibration films zero reference p...

Page 20: ...max 10 m on the free field Temperature range Storage 10 C to 60 C 14 F to 140 F Operation 0 C to 50 C 32 F to 122 Power supply 2 x batteries AA 1 5V alkaline or 2 x rechargeable batteries AA 1 2V Dime...

Page 21: ...e pole and negative pole to avoid short circuiting the device Do not use new and old batteries or batteries of a different kind together This could lead to miss functions in your device Remove empty b...

Reviews: