background image

Instruction Manual QNix

®

8500 

- 17 - 

first 4 digits indicate the software version without the decimal dot (so “014c” means “1.4c”), 
and the last 3 digits show the hardware revision of the gauge, also without the decimal dot. 

10.4 

I cannot activate the menu 

Use the 

▼ button. If 2 or more memory batches are defined and the memory mode is select-

ed, the 

◄ and ► buttons select the next/previous batch, and the ▼ button switches to menu 

entry (see also 8.3.5.1). 

11 

Installing the

 

PC Software 

The PC-software is automatically installed after inserting the CD into your PCs CD-ROM 
drive. 

12 

Installing the USB-RF-Dongle Drivers 

You find the drivers for the USB-RF-dongle on the

 

PC-software CD.  

After the dongle has been connected to the PC for the first time, the operating system of your 
PC will open a window showing a message that new hardware has been found and that nec-
essary drivers have to be installed. Enter the path of the drive holding the CD in the desig-
nated field or use the “Search” function. Then start the installation.  
This installation process is necessary only the first time the dongle is used. Please also refer 
to the installation guide on the CD.  

13 

Setting the Gauge Options with the PC-Software 

The setup options described in § 6can also be programmed and then downloaded into the 
gauge using the PC-software. This is especially useful when entering the names for calibra-
tions and batches.   

To do this, connect the RF-dongle to your PC and start the software. Also, activate the wire-
less interface of the gauge (see § 8.5.2). 

After the PC-software identified the gauge, the possible menu options and their settings are 
displayed in the software menu “Gauge Settings” or can be loaded from the gauge by click-
ing “Load Settings from Gauge”. 
You can now enter and modify all gauge settings on your PCs monitor. Only after all inputs 
are finis

hed the settings are downloaded into the gauge by clicking on “Transfer Settings to 

Gauge”.   

14 

Loading a new Language into the Gauge  

Use the "LanguageEditor" of the PC-software to load a new language into the gauge. 

The LanguageEditor contains a chart showing all commands and messages appearing on 
the display.  

15 

Updating the Software in the Gauge and Probe 

The software in the gauge and probe can be updated, if required. Enhancements of the soft-
ware are available for download on the internet sit

www.automation.de

 

After downloading to the PC the updates can transferred to the gauge and probe with the 
RF-interface. 

Summary of Contents for QNix 8500

Page 1: ......

Page 2: ...Instruction Manual QNix 8500 2...

Page 3: ...Instruction Manual QNix 8500 3...

Page 4: ...Averaging 11 8 3 Data Memory 11 8 3 1 Delete All 12 8 3 2 Delete values 12 8 3 3 Data Memory Mode ON OFF 12 8 3 4 Creating a New Batch 12 8 3 5 Selecting Editing a Batch 12 8 4 Statistics 13 8 4 1 Se...

Page 5: ...e PC Software 17 14 Loading a new Language into the Gauge 17 15 Updating the Software in the Gauge and Probe 18 16 Display Symbols 18 17 Scope of Delivery 19 18 Technical Data 20 Notice This instructi...

Page 6: ...to strong magnetic fields Remove the batteries from the gauge and store them separately if the device is not used for a longer period The gauge is not waterproof Do not expose it to water 3 Changing...

Page 7: ...ring head Within the measuring range of the probe measurements like e g distance measurements are possible that way 6 Setting the Zero Reference When using the gauge for the first time after changing...

Page 8: ...u For your convenience find a diagram of the menu structure showing all possible menu op tions on the first two pages of this instruction manual Depending on the version of your gauge only a part of t...

Page 9: ...d calibrations under the menu item Define New Calibration To select a certain calibration use the and buttons Subse quently pressing opens the corresponding options menu A dot at the right side of the...

Page 10: ...a number of zero reference readings max 19 to be taken with the probe before the gauge will calculate the average zero reference based on the readings taken After selecting the menu item pressing the...

Page 11: ...t To set or delete the lower limit please follow the instructions given under 8 2 3 Readings below the lower limit are indicated in the display with the symbol and with an au dio warning signal 8 2 5...

Page 12: ...firm it by pressing For corrections of prior digits simply press the backspace button at any time NOTE When creating a new batch after other batches i e B 0001 B 0002 have been de leted the gauge stil...

Page 13: ...he statistics functions are calculated per memory block The gauge determines the following statistical parameters average value standard deviation maximum minimum Within the statistics menu two differ...

Page 14: ...ing of the display To change the display orientation press the button then select the setting Normal or Flip using and and finally confirm by pressing again NOTE When the display is flipped the functi...

Page 15: ...he and buttons Confirm with For corrections use the backspace button 8 5 6 2 Setting the Date Set the date the same way you set the time 8 5 7 The button The button is used to set the zero reference s...

Page 16: ...coating you want to measure If an uncoated piece is unobtainable you can use the standard plates that are supplied with your gauge Refer to 6 for further details Please note that it is the probe that...

Page 17: ...tion guide on the CD 13 Setting the Gauge Options with the PC Software The setup options described in 6can also be programmed and then downloaded into the gauge using the PC software This is especiall...

Page 18: ...rror operating error Infi readings outside measuring range wrong type of surface readings exceeding upper limit readings below lower limit x average standard deviation wireless interface active p radi...

Page 19: ...th clip Plastic case for transportation and storage Optional Accessories Fe probe 2 mm Fe probe 5 mm NFe probe 2 mm Dual probe Fe NFe 2 2 mm Dual probe Fe NFe 5 2 mm Calibration films zero reference p...

Page 20: ...max 10 m on the free field Temperature range Storage 10 C to 60 C 14 F to 140 F Operation 0 C to 50 C 32 F to 122 Power supply 2 x batteries AA 1 5V alkaline or 2 x rechargeable batteries AA 1 2V Dime...

Page 21: ...e pole and negative pole to avoid short circuiting the device Do not use new and old batteries or batteries of a different kind together This could lead to miss functions in your device Remove empty b...

Reviews: