background image

further determined with the volt/ohmmeter).  If you find 

the break, resolder the connection, wrap the inner wires 

in separate pieces of electrical tape, retest the valve, and 

reassemble the sheath following the above instructions in 

reverse order.

Valve Returns & Replacing a Valve

If you need to return a defective valve, you may either return the 

entire valve assembly for repair / evaluation, or remove and return the 

individual valve from its case.  This only involves five screws:  remove the 

two outer plate screws holding all of the valves into the black metal case, 

the single screw holding the plastic ‘sheath’ into its underside hole, the 

electrical tape at the top and bottom of the sheath, the offending valve’s 

cable from the plug end up through the valve casing, and finally, the two 

screws holding the valve on to the front plate.

Please package the valve(s) carefully and ship freight prepaid to AutoMate 

Scientific’s address on the back cover of this manual.  No RMA number is 

needed as long as you include a note with the following:

1)  Your name, company/institution name, phone number, and an address 

to return the repaired valve to.

2)  Your original invoice number and date, if available.

3)  A sentence stating the nature of the problem and any steps you have 

already taken to fix or identify it.  Does the valve energize / click?  De-

energize?  Open / close?  Leak?

4)  Whether the valve was driven by any device other than an AutoMate 

Scientific controller (and its operating and holding voltages).

5)  The application used with the valve, along with the exact chemical(s) 

and the duty cycle (pulse) rate if used.

6)  A Statement of Chemical Exposure as follows: 

a) These valves have not been exposed to chemicals other than H

2

except as described above.  b) All applicable Safety and Handling data 

sheets for the above described chemicals are enclosed.  c) We accept 

full responsibility for any injury to AutoMate Scientific employees or 

employees of the valves original manufacturer caused by handling of 

the residues of chemicals contained in or on the valves to be returned.

7)  An authorized signature, title, and date.

AutoMate Scientific will gladly fax you a form with easy blanks to enter 

the above information.  We will process your return and notify you of 

further information as quickly as possible.

3

Heat 

shrink 

on each 

cable

Leaks and Replacing Valve Fittings

The second bane of all liquid delivery apparatus!  These hints may help:

1)  Make sure the valve is turned off! (as opposed to leaking)

2)  See all of the guidelines in the previous “Faulty Valves” section -  

especially #1.

3)  Identify whether the valve leaks constantly, under pressure, inside or 

outside.  Try backflushing if there is an internal leak.

4)  Check and replace obvious loose or worn tubing.

5)  Outside leaks usually occur at the fittings and can be stopped with a bit 

of silicone glue, epoxy, super glue (Cyanoacrylate) or PTFE tape. Apply 

sealant around the threads of the dry fitting and screw back into the 

valve.  Be sure not to use too much glue or tape, as they can easily cause 

clogs or leaks of their own.  Overtightening can also cause the plastic 

fittings to leak – hand tighten.

Glue on valve fittings is intended as a gap filler for small openings between 

the threads and hole, not as an adhesive.  Few glues will adhere to PTFE.

Hardware Troubleshooting

Should the ValveLink8.2's LEDs fail to light while plugged-in, check the 

power connections.  Disconnect all valves to check for a short.  Remember 

that ValveLink8.2 valve LEDs do not light if valves are not plugged-in.  

The ValveLink8.2 uses an internal "poly" fuse which will reset itself after an 

overload situation has been fixed.  Try unplugging the ValveLink8.2 for a 

while then reconnecting it.

39

38

Summary of Contents for ValveLink 8.2

Page 1: ...ValveLink8 2 User s Manual AutoMate Scientific R E A D Y F O R R E S E A R C H TM...

Page 2: ...ve Tubing Valve Care Perfusion Pencil Flow Rates Perfusion Pencil Multi barrel Manifold Perfusion Pressure Kit Josef Kewekordes II Heath Lukatch Dave Blau ValveLink8 2 Part 01 18 For indoor use only f...

Page 3: ...RCA jacks on the back of the box Why not connect valves directly to your computer Several reasons 1 Most computer interfaces provide 5V signals not strong enough to drive most valves 6 to 12V DC and h...

Page 4: ...wheels and pipette pullers ValveLink8 2 Laboratory Applications Perfusion A ValveLink8 2 can be programmed to precisely switch solutions perfusing over biological samples for physiological research T...

Page 5: ...on inline heater plus multi channel focal drug delivery maintain bath temperature and rapid drug wash out with a high flow bath line while quickly switching 4 8 preheated solutions through the Perfusi...

Page 6: ...ld also see the photo on page 5 Find additional photos and diagrams of the pressure system on page 15 2 3 4 1 Assemble ringstand base and rod 2 Attach valve unit 3 Insert syringes with stopcocks Pinch...

Page 7: ...ine directly to the outflow of the manifold for running cleaning solutions and drying your lines quickly Determine dead volume from the manifold to prep by measuring time and volume required to flush...

Page 8: ...e Tips reservoir height 40cm Gravity 0 2psi air pressure 0 07ml min 4 6ml hour 8psi air pressure 0 25ml min 15ml hour PTFE Micro manifold Use To use the micro manifold you either need pinch valves or...

Page 9: ...and removable tips are shipped non sterile They can be chemically sterilized or autoclaved However repeated autoclaving may weaken the adhesive bond between the luer lock fitting and the needle Maint...

Page 10: ...es mirroring the computer signals low off high on with one millisecond accuracy Usually data acquisition software like pClamp Pulse LabView etc or custom user software is programmed to send digital si...

Page 11: ...nt panel analog output BNC as an event marker every time a valve opens or closes The marker voltage occurs regardless of how the valve was opened or closed i e manual pushbutton or computer digital in...

Page 12: ...an electrical path for the spill sensor circuitry In this way a drop of liquid passing the sensor strip appears as a sharp drop in electrical resistance between the two otherwise disconnected copper c...

Page 13: ...traces A similar conductive epoxy is also available It makes the connection of two wires to your silver traces even easier Simply draw two parallel silver lines approximately 1 16 1 mm apart on any s...

Page 14: ...between two ValveLink8 2s and control up to 16 valves with a single 4 bit TTL source Set the first ValveLink8 2 to Mode 2 set the second to Mode 3 and connect TTL input bits 1 4 from both ValveLink8 2...

Page 15: ...ur version of Windows the first time a ValveLink8 2 is installed by USB cable Windows will find new hardware A dialog box will appear asking if you want Windows to connect to Windows Update to search...

Page 16: ...on rotary switch on the back of each ValveLink8 2 next to its USB port If you connect multiple ValveLink8 2s to the computer be sure to set each one to a different address If you connect a ValveLink8...

Page 17: ...one valve will be open at a time on all of the networked controllers When the System Status switch is set to Local Control you can click the Mode button on any ValveLink8 2 sub window to access a dia...

Page 18: ...ar as BNC connectors on the front panel Digital outputs 4 7 are located on the back panel of newer DigiData interfaces Older DigiData 1200 s only have four outputs 2 The Modes chapter explains the two...

Page 19: ...tubing at an angle and wetting it makes it easier to insert Backflow When a manifold is used in a plumbing arrangement none of its ports can be left unconnected or liquid will simply flow backwards o...

Page 20: ...g see 1 above and the section below If it qualifies for all three of the above AutoMate Scientific will accept the valve for subsequent return to the factory Repair or replacement is at their discreti...

Page 21: ...e Safety and Handling data sheets for the above described chemicals are enclosed c We accept full responsibility for any injury to AutoMate Scientific employees or employees of the valves original man...

Page 22: ...e power requirements of this product 6 Do not allow anything to rest on the power cord Do not locate this product where people will walk on the cord 7 If you have to use an extension cord with this pr...

Page 23: ...nsor licensees be liable for contingent special direct indirect or consequential damages for the breach of any express or implied warranty or resulting from the use failure or malfunction of any produ...

Page 24: ...entific Inc 812 Page St Berkeley CA 94710 U S A 800 998 MATE 6283 international 510 845 6283 fax 510 280 3795 e mail info autom8 com www http www autom8 com AutoMate Scientific R E A D Y F O R R E S E...

Reviews: