3-way Valves

The ValveLink8.2 also supports 3-way, single-solenoid valves (2 wires 

each).  Operate the ValveLink8.2 as you normally would for 2-way valves, 

after determining which of two inlets or outlets are opened with the on and 

off states.  Note that one inlet will always be open with these valves.

Power Supply

The ValveLink8.2 is normally supplied with a 1.5 amp wall transformer.  

This is adequate for any valves provided by AutoMate Scientific or a 

total of 1 amp of valves opened simultaneously.  Please contact AutoMate 

Scientific to request a larger power supply if you wish to exceed 1 amp of 

valve power (i.e. two 12 watt valves opened at once).

Case and Valve Grounding

This equipment must be earth grounded.  Use a #6 screw in any hole 

on either side of the ValveLink8.2 case for grounding.  Order AutoMate 

Scientific part no. 01-05 “Low noise, valve and case grounding package” 

for an extra grounding wire attached to each individual valve extending 

back to the controller.  Each of these items grounds four (4) valves.  Please 

order two for eight channel perfusion systems.  This item must be ordered 

at the same time as the valves.  Connect the ground wire from the valves to 

a screw on the side of the controller case or to your rig's ground point.

chamber where liquid would pass if it overflows.  It can also be wrapped 

around the outside of a light condenser or inverted microscope objective 

turret where liquid would flow if it drips out of the perfusion chamber.

Researchers may create or use their own spill sensors with the 

ValveLink8.2.  Conductive mats or other commercially-available spill 

sensors generally use the same principle as AutoMate Scientific's copper 

strips.  Likewise, you can easily create your own custom sensor using two 

bare wires or conductive pens or epoxy.  Inexpensive conductive pens 

like those from "Circuit Works

" are designed for repairing circuit boards.  

They leave lines which quickly dry into silver traces.  A similar conductive 

epoxy is also available.  It makes the connection of two wires to your silver 

traces even easier.  Simply draw two parallel silver lines approximately 

1

/

16

" (1 mm) apart on any surface where leaking liquids would pass.  Make 

sure the lines do not touch, or draw one wide line and etch a gap down the 

middle once it has dried.  Then connect a small wire from each trace back 

to the spill sensor port on the ValveLink8.2.

Another nice feature of the ValveLink8.2 spill sensor circuit is its dynamic 

recalibration.  Competing spill sensors simply watch for a low resistance 

and trigger.  Unfortunately, their conductive mats keep a low resistance 

and cannot be reused until they dry.  The ValveLink8.2 lets you press 

any button to override the spill.  Because it is microprocessor-based, the 

controller keeps a running value for the sensor resistance and watches for 

a 5% drop in resistance.  Therefore, even a damp sensor can be used again 

immediately as the controller adjusts to the new resistance and watches for 

any subsequent leaks by another proportional resistance drop.  As the mat 

dries, the ValveLink8.2 is constantly monitoring the increasing resistance 

and watching for a drop from the latest value.

Where is the power switch?  Sleep

The ValveLink8.2 does not have a power switch.  As long as it is plugged-

in, it is on.  The ValveLink8.2 is perfectly happy being left on forever 

and consumes very little power when all valves are closed.  However, an 

optional sleep mode has been included.  Press and hold the "Mode/Sleep" 

button until the blue and yellow Mode LEDs begin flashing (about three 

seconds) to put the ValveLink8.2 to sleep.  In this state all valves and valve 

LEDs are off, and all inputs (buttons, TTLs, USB, etc.) are ignored.  Press 

the Mode button again to wake the controller.  Sleep mode can also be used 

as an emergency "off" switch if you need to quickly close all valves and 

stop the computer inputs.

12V AC Power (up to 2.4 amps @ 50-60Hz)

USB Port and address

(see USB chapter)

Digital inputs

(see earlier section)

Back Panel

12 volt DC valves

1A (12W) max per channel

2A (24W) max total

Spill

sensor

(prev. pages)

23

22

Summary of Contents for ValveLink 8.2

Page 1: ...ValveLink8 2 User s Manual AutoMate Scientific R E A D Y F O R R E S E A R C H TM...

Page 2: ...ve Tubing Valve Care Perfusion Pencil Flow Rates Perfusion Pencil Multi barrel Manifold Perfusion Pressure Kit Josef Kewekordes II Heath Lukatch Dave Blau ValveLink8 2 Part 01 18 For indoor use only f...

Page 3: ...RCA jacks on the back of the box Why not connect valves directly to your computer Several reasons 1 Most computer interfaces provide 5V signals not strong enough to drive most valves 6 to 12V DC and h...

Page 4: ...wheels and pipette pullers ValveLink8 2 Laboratory Applications Perfusion A ValveLink8 2 can be programmed to precisely switch solutions perfusing over biological samples for physiological research T...

Page 5: ...on inline heater plus multi channel focal drug delivery maintain bath temperature and rapid drug wash out with a high flow bath line while quickly switching 4 8 preheated solutions through the Perfusi...

Page 6: ...ld also see the photo on page 5 Find additional photos and diagrams of the pressure system on page 15 2 3 4 1 Assemble ringstand base and rod 2 Attach valve unit 3 Insert syringes with stopcocks Pinch...

Page 7: ...ine directly to the outflow of the manifold for running cleaning solutions and drying your lines quickly Determine dead volume from the manifold to prep by measuring time and volume required to flush...

Page 8: ...e Tips reservoir height 40cm Gravity 0 2psi air pressure 0 07ml min 4 6ml hour 8psi air pressure 0 25ml min 15ml hour PTFE Micro manifold Use To use the micro manifold you either need pinch valves or...

Page 9: ...and removable tips are shipped non sterile They can be chemically sterilized or autoclaved However repeated autoclaving may weaken the adhesive bond between the luer lock fitting and the needle Maint...

Page 10: ...es mirroring the computer signals low off high on with one millisecond accuracy Usually data acquisition software like pClamp Pulse LabView etc or custom user software is programmed to send digital si...

Page 11: ...nt panel analog output BNC as an event marker every time a valve opens or closes The marker voltage occurs regardless of how the valve was opened or closed i e manual pushbutton or computer digital in...

Page 12: ...an electrical path for the spill sensor circuitry In this way a drop of liquid passing the sensor strip appears as a sharp drop in electrical resistance between the two otherwise disconnected copper c...

Page 13: ...traces A similar conductive epoxy is also available It makes the connection of two wires to your silver traces even easier Simply draw two parallel silver lines approximately 1 16 1 mm apart on any s...

Page 14: ...between two ValveLink8 2s and control up to 16 valves with a single 4 bit TTL source Set the first ValveLink8 2 to Mode 2 set the second to Mode 3 and connect TTL input bits 1 4 from both ValveLink8 2...

Page 15: ...ur version of Windows the first time a ValveLink8 2 is installed by USB cable Windows will find new hardware A dialog box will appear asking if you want Windows to connect to Windows Update to search...

Page 16: ...on rotary switch on the back of each ValveLink8 2 next to its USB port If you connect multiple ValveLink8 2s to the computer be sure to set each one to a different address If you connect a ValveLink8...

Page 17: ...one valve will be open at a time on all of the networked controllers When the System Status switch is set to Local Control you can click the Mode button on any ValveLink8 2 sub window to access a dia...

Page 18: ...ar as BNC connectors on the front panel Digital outputs 4 7 are located on the back panel of newer DigiData interfaces Older DigiData 1200 s only have four outputs 2 The Modes chapter explains the two...

Page 19: ...tubing at an angle and wetting it makes it easier to insert Backflow When a manifold is used in a plumbing arrangement none of its ports can be left unconnected or liquid will simply flow backwards o...

Page 20: ...g see 1 above and the section below If it qualifies for all three of the above AutoMate Scientific will accept the valve for subsequent return to the factory Repair or replacement is at their discreti...

Page 21: ...e Safety and Handling data sheets for the above described chemicals are enclosed c We accept full responsibility for any injury to AutoMate Scientific employees or employees of the valves original man...

Page 22: ...e power requirements of this product 6 Do not allow anything to rest on the power cord Do not locate this product where people will walk on the cord 7 If you have to use an extension cord with this pr...

Page 23: ...nsor licensees be liable for contingent special direct indirect or consequential damages for the breach of any express or implied warranty or resulting from the use failure or malfunction of any produ...

Page 24: ...entific Inc 812 Page St Berkeley CA 94710 U S A 800 998 MATE 6283 international 510 845 6283 fax 510 280 3795 e mail info autom8 com www http www autom8 com AutoMate Scientific R E A D Y F O R R E S E...

Reviews: