Faulty Valves

The bane of all liquid delivery apparatus!  Despite AutoMate Scientific’s 

best efforts and pursuit of high quality valve manufacturers, some 

valves fail.  Space shuttle launches have often been delayed due to valve 

problems.  We can only pass along our valve sup plier’s one year warranty 

with their disclaimer “Improper use or mishandling of the units, in the 

opinion of the manufacturer, voids this warranty!” (their exclamation mark 

too.)  Please request a copy of AutoMate Scientific's "Valve Troubleshooting 

Guide" to diagnose and sometimes repair valve problems.  In order to 

minimize valve trouble, please observe these guidelines:

1)  Clean your valves after each use.  Accumulated debris or precipitate 

will rapidly cause leaking valves and seals.  Use an in-line filter if you 

expect a chance of particles in your solutions.  Put your AutoMate 

Scientific controller’s programmability to good use by writing a washing 

protocol.  At the end of the day, fill each reservoir with cleaning solution 

(distilled water or appropriate liquid), and open all valves.  A protocol can 

be easily written to flush each valve in sequence.  Saline solutions allowed 

to dry in stainless steel valves will cause them to rust.  This is considered 

improper use.  We also recommend rinsing your valves once before 

their first use.

Beware of flaking PTFE tape or silicone fitting glue.  Large debris are often 

the cause of valve failure.  Try backflushing a valve that is suspected to 

have failed for this reason by connecting a tube and syringe to the outflow 

of the valve and pushing liquid while it is energized.  Sometimes this 

ejects the clog.

2)  Do not operate AutoMate Scientific valves with homemade controllers 

delivering over 12V DC, or attempt to cycle them rapidly (< 1ms).

3)  Do not exceed the rated operating pressure (> 30psi).

Before returning any valves to us please check:

1)  That it was purchased in the last 12 months (AutoMate Scientific and 

valve manufacturers record each by serial number.)

2)  That the valve is faulty and not the controller channel – find a valve that 

works in a particular output channel and try the bad valve in that port 

to see that it still does not function.

3)  If it is leaking, see #1 above and the section below.

If it qualifies for all three of the above, AutoMate Scientific will accept the 

valve for subsequent return to the factory.  Repair or replacement is at 

their discretion, as AutoMate Scientific does not have the ability to open 

the valves to determine the cause of trouble.  We may be able to loan 

replacement valves, which may be billed if the valve factory finds abuse on 

the user’s part and refuses to honor the warranty.

Stuck Pinch Valves

Occasionally pinch valves will stick closed, especially after they have 

gotten wet.  Usually this is a sign that they are about to completely fail, 

but sometimes they can be rejuvenated by simply prying them open or 

lubricating them.  First energize (turn on) the valve; then, using a flat 

screwdriver, gently force the white plunger open where it normally 

pinches the tube closed.  The valve may begin functioning normally.  Also, 

pinch valves will not usually work without a piece of pinch tubing 

installed!  You can also remove the pinch tubing and spray a drop of WD-

40 lubricant or isopropyl alcohol down the pinch valve plunger.  Replace 

the tubing.  This will often add more life to a dying pinch valve.

Valve Cables

One problem that is easy to diagnose and fix is a broken cable connection.  

Use any ohmmeter to measure the resistance across the inner pin and 

outer ‘barrel’ of the RCA plug end of the valve cable (while unplugged 

from the controller.)  A resistance of zero ohms indicates a short, or infinite 

resistance means one of the wires is broken in the cable assembly.  This 

occurs most often at the joint between the heavy black cable and the valve 

lead wires – under the heat-shrink tubing, inside the outer black sheath, 

a few inches from the valve itself.  This can often be reached without 

completely disassembling the valve enclosure.

Simply 1) remove the screw inside the black valve box holding the plastic 

sheath in its hole,  2) remove the black tape at the top of the sheath and 

peel it back a couple of inches, find the offending cable (trace it back from 

the bad valve), and 3) carefully slice its heat-shrink tubing longitudinally 

to remove it.  The wires may 

have come apart or broken 

here, or inside the next piece of 

electrical tape isolating the two 

leads, otherwise within the cable 

or valve itself (which can be 

1

2

Valve enclosure (shown without front plate)

Plastic ‘split loom’ cable sheath.

37

36

Summary of Contents for ValveLink 8.2

Page 1: ...ValveLink8 2 User s Manual AutoMate Scientific R E A D Y F O R R E S E A R C H TM...

Page 2: ...ve Tubing Valve Care Perfusion Pencil Flow Rates Perfusion Pencil Multi barrel Manifold Perfusion Pressure Kit Josef Kewekordes II Heath Lukatch Dave Blau ValveLink8 2 Part 01 18 For indoor use only f...

Page 3: ...RCA jacks on the back of the box Why not connect valves directly to your computer Several reasons 1 Most computer interfaces provide 5V signals not strong enough to drive most valves 6 to 12V DC and h...

Page 4: ...wheels and pipette pullers ValveLink8 2 Laboratory Applications Perfusion A ValveLink8 2 can be programmed to precisely switch solutions perfusing over biological samples for physiological research T...

Page 5: ...on inline heater plus multi channel focal drug delivery maintain bath temperature and rapid drug wash out with a high flow bath line while quickly switching 4 8 preheated solutions through the Perfusi...

Page 6: ...ld also see the photo on page 5 Find additional photos and diagrams of the pressure system on page 15 2 3 4 1 Assemble ringstand base and rod 2 Attach valve unit 3 Insert syringes with stopcocks Pinch...

Page 7: ...ine directly to the outflow of the manifold for running cleaning solutions and drying your lines quickly Determine dead volume from the manifold to prep by measuring time and volume required to flush...

Page 8: ...e Tips reservoir height 40cm Gravity 0 2psi air pressure 0 07ml min 4 6ml hour 8psi air pressure 0 25ml min 15ml hour PTFE Micro manifold Use To use the micro manifold you either need pinch valves or...

Page 9: ...and removable tips are shipped non sterile They can be chemically sterilized or autoclaved However repeated autoclaving may weaken the adhesive bond between the luer lock fitting and the needle Maint...

Page 10: ...es mirroring the computer signals low off high on with one millisecond accuracy Usually data acquisition software like pClamp Pulse LabView etc or custom user software is programmed to send digital si...

Page 11: ...nt panel analog output BNC as an event marker every time a valve opens or closes The marker voltage occurs regardless of how the valve was opened or closed i e manual pushbutton or computer digital in...

Page 12: ...an electrical path for the spill sensor circuitry In this way a drop of liquid passing the sensor strip appears as a sharp drop in electrical resistance between the two otherwise disconnected copper c...

Page 13: ...traces A similar conductive epoxy is also available It makes the connection of two wires to your silver traces even easier Simply draw two parallel silver lines approximately 1 16 1 mm apart on any s...

Page 14: ...between two ValveLink8 2s and control up to 16 valves with a single 4 bit TTL source Set the first ValveLink8 2 to Mode 2 set the second to Mode 3 and connect TTL input bits 1 4 from both ValveLink8 2...

Page 15: ...ur version of Windows the first time a ValveLink8 2 is installed by USB cable Windows will find new hardware A dialog box will appear asking if you want Windows to connect to Windows Update to search...

Page 16: ...on rotary switch on the back of each ValveLink8 2 next to its USB port If you connect multiple ValveLink8 2s to the computer be sure to set each one to a different address If you connect a ValveLink8...

Page 17: ...one valve will be open at a time on all of the networked controllers When the System Status switch is set to Local Control you can click the Mode button on any ValveLink8 2 sub window to access a dia...

Page 18: ...ar as BNC connectors on the front panel Digital outputs 4 7 are located on the back panel of newer DigiData interfaces Older DigiData 1200 s only have four outputs 2 The Modes chapter explains the two...

Page 19: ...tubing at an angle and wetting it makes it easier to insert Backflow When a manifold is used in a plumbing arrangement none of its ports can be left unconnected or liquid will simply flow backwards o...

Page 20: ...g see 1 above and the section below If it qualifies for all three of the above AutoMate Scientific will accept the valve for subsequent return to the factory Repair or replacement is at their discreti...

Page 21: ...e Safety and Handling data sheets for the above described chemicals are enclosed c We accept full responsibility for any injury to AutoMate Scientific employees or employees of the valves original man...

Page 22: ...e power requirements of this product 6 Do not allow anything to rest on the power cord Do not locate this product where people will walk on the cord 7 If you have to use an extension cord with this pr...

Page 23: ...nsor licensees be liable for contingent special direct indirect or consequential damages for the breach of any express or implied warranty or resulting from the use failure or malfunction of any produ...

Page 24: ...entific Inc 812 Page St Berkeley CA 94710 U S A 800 998 MATE 6283 international 510 845 6283 fax 510 280 3795 e mail info autom8 com www http www autom8 com AutoMate Scientific R E A D Y F O R R E S E...

Reviews: