102
Bandwidth
20MHz
Accuracy
Voltage: 1%; Time: 50ppm
Sensitivity
10mV/div to 20V/div
Input Ranges (full scale)
±50mV to ±100V in 11 ranges
Input Impedance
1MΩ in parallel with 22pF
Input Type
Single-ended, BNC connector
Input Coupling
Software selectable AC/DC
Overload Protection
±200V on single input
Max. Sampling Rate (Single
Shot)
1 or 2 channels in use
3 or 4 channels in use
80MS/s*
20MS/s
Buffer Memory
32M samples shared among active channels
Waveform Buffer
Up to 1000 waveforms
Timebase Ranges
100ns/div to 1000s/div
Advanced Features
Math channels, Measurements
Triggers
Description
Source
Any input channel
Basic Triggers
Auto, Normal, Single, None
Advanced Triggers
Rising edge, Falling edge
Environmental
Description
Operating Temp. Range
Storage Temp. Range
Storage Humidity Range
0 to 50°C (15 to 40°C for quoted accuracy)
-20 to +60°C
5 to 95%RH, Non-condensing
Physical Characteristics
Description
Dimensions
(Protection
rubber case included)
190x115x38 mm
Weight
< 0.5 kg
General
Description
PC Interface
USB 2.0
– cable supplied
Power Requirements
Powered from USB port