Mechanical settings
US
-
2000
(
DCP 2000 V
V5)
SERVICE MANUAL
CE
Fertigstellung: 02.16
US - 2000
g
g
Serviceanleitung_US2000V5_EN_V0.0.docx
Page 11 of 23
You must not exchange sonotrode and generator separately! Exchange both when either fails. Warranty
is void when only one part has been replaced.
Preparation:
1. The lower ferrite core has to be flush with the end face of the
white screw. You’ll find it on the bottom of the ultrasonic
generator.
Connecting the ampere meter:
1.
Disconnect the power
plug from the machine.
2. Unplug the ultrasonic generator plug and unscrew the phase
line (brown) from the plug.
3. Connect the phase line to the analog ampere meter.*
4. Connect the other terminal of the ampere meter to the
phase terminal of the ultrasonic generator.
5. Make sure the wires are connected properly and switch on
the machine.
*Do not use other ampere meters than the ampere meter distributed
by ATS (
ʋ
120526). We can’t guarantee that the values in this
document are equal to the values displayed by other measuring
equipment.
Adjusting the 25mm sonotrode & generator set:
1. Set the welding time to 300ms.
2. Turn the white screw clockwise or counterclockwise to adjust the current consumption of the generator
during welding.
Adjust the current to 4.0 Amperes.
3. Fix the white screw with the lock nut on it.
4. Reduce the welding time to have an accurate welding (ideal welding time depends on the material used).
The sonotrode’s noise should be a high tone and it should not “grumble”. If it does, reduce the current
consumption.