background image

     

9

LARGE DIAMETER WELLS: 

 Dug or bored wells should be disinfected in the same way as a drilled 

one.  Lower the water level as much as possible, remove the sand, silt and debris, and then treat with the chlorine 

solution.  Mix thoroughly by circulating the water back into the well  and use the hose to rinse the interior  

lining of the well.  Do not try to disinfect an unprotected, unlined well because new seepage or surface  

contamination will flow into the water about as fast as you can disinfect it.  Disinfect the pipeline distribution 

system as indicated for drilled wells.

SPRINGS AND CISTERNS

:

  Mix about ½ cup of household bleach in 5 gallons of water and use this to 

scrub the walls of the spring box or holding tank.  With a constant flow of fresh water from the spring, there is  

probably no way of detaining the chlorine solution in the reservoir for more than a few minutes.  However, the 

chlorinated water should flow through the pipeline to disinfect the distribution system.  Cisterns can be disin-

fected in the same way but a source of clean water will be needed to flush the dirty waste out of the system.

For additional information about how to protect wells and springs and keep them from becoming contaminated, 

call or visit your local Cooperative Extension office, or your nearest certified water treatment specialist.

TABLE 1

 **  

Quantity of solution mixed - 5.25% available chlorine (laundry bleach) for disinfecting   

 

 

wells, or 52,500 P.P.M..

WQA recommends 50mg/l or ppm chlorine concentration.

  
Formula -   C

X V

2

 / C

1

 = V

1

 

C

1

= Household Bleach (52,500 P.P.M.) 

 V

1

= Chlorine Amount Needed

 C

2

= 50 mg/L  V

2

= 80 gallons holding time

 

I.E.  50-ml/g X 80 gal= 4000/52,500= .08 gallons of chlorine (5.25%)

 

       .08 gal chlorine (5.35%) X 128 (oz/gal) = 10.24 oz (5.25%)

 

Dug Wells - 3 to 4 feet diameter - 4 cups per foot of water

 

Drilled Wells - 3 to 8 inch diameter - 1 cup per foot of water

 

TOO MUCH CHLORINE IS BETTER THAN TOO LITTLE:

** 

In situations where it is inconvenient to determine depth of water or diameter of a drilled well, a  

minimum of 

1/2

 gallon of household bleach may be used for wells up to 8 inches in diameter with estimated to 

be less than 80 feet deep; 1 gallon should be used for similar size wells with water deeper than 80 feet.  In case 

of a well yielding more than 50 gallons per minute, special procedures are required.  Seek the advice of a  

certified water treatment specialist.

Wait a day or two before you have another sample tested. 

 Do not take a sample for testing if the odor of  

chlorine is still present in water.

Summary of Contents for Safe Lite SL-2V

Page 1: ...Models SL 2V 8V It is important that those responsible for the installation of this equipment as well as the owner operator read this manual and carefully follow the instructions and guidelines Revis...

Page 2: ...r sold by the manufacturer distributor may cause an unsafe condition 7 Intended for indoor use Do not install this disinfection system where it will be exposed to the weather Do not store this disinfe...

Page 3: ...Flow rate must not exceed rated capacity of the unit DESCRIPTION OF EQUIPMENT The Safe Lite series are of unique design with an ultraviolet germicidal lamp housed within a single quartz sleeve surroun...

Page 4: ...personnel look at a U V lamp in operation EXCEPT through an external Site Port lens located on the outside of the unit 3 U V disinfection units must always be properly grounded INSTALLATION The Safe L...

Page 5: ...selected Simply hand press or slightly tap in the flow control until it sits on the inside ledge of the machined port 3 Using the mounting box provided secure unit to wall or other surface Make sure t...

Page 6: ...eposit onto the quartz sleeve After sufficient film has formed on the quartz sleeve the ability of the ultraviolet germicidal rays to pass though the quartz sleeve and into the water may be impaired T...

Page 7: ...shortwave U V to effectively kill bacteria Failure to replace the U V lamps every 9 000 hours may cause bacteriological breakthrough Should the use of the disinfection unit be intermittent in no case...

Page 8: ...o determine how much chlorine solution should be used In some instances removing the seal to measure the water can be a difficult task and it is easier to guess at the quantity of disinfectant needed...

Page 9: ...w to protect wells and springs and keep them from becoming contaminated call or visit your local Cooperative Extension office or your nearest certified water treatment specialist TABLE 1 Quantity of s...

Page 10: ...ater System Using Well Sanitizer Pellets Table 1 Well Diameter Inches Weight of Pellets lbs oz Cups of Pellets Number of Pellets To produce a 400 P P M chlorine dosage NOTE Pellets Weight 1 14 gram ea...

Page 11: ...n water back down the well this also rinses upper portion of well After about 15 minutes of recirculation the water a strong chlorine odor should be apparent Turn off hose 6 Bypass water softener and...

Page 12: ...gulator or above 120 volts Line cord disconnected Check replace or outlet defective Defective U V lamp Replace Defective lamp ballast Check output voltage Replace ballast Loose open circuit wire Trace...

Page 13: ...proval by Aqua Treatment Service Inc the Dealer Distributor should return the item to Aqua Treatment Service Inc 194 Hempt Road Mechanicsburg PA 17050 prepaid If a part proves to be defective it will...

Page 14: ...120V 03 03 AMPS 120V 25 25 INLET OUTLET SIZE 3 8 STD MNPT 3 4 STD MNPT TOTAL DIMENSION INCL WALL MOUNT L X W X D 15 5 X 5 X 5 21 X 5 X 5 DISINFECTION CHAMBER MATERIAL WETTED PARTS 304 SS TUBING 304 SS...

Page 15: ...bottle or sterile container being used Holding the cap in a down position the operator should then hold his breath while taking the sample so as to avoid oral contamination of the sample The operator...

Page 16: ...16 WIRING DIAGRAM...

Page 17: ...17...

Page 18: ...18...

Page 19: ...19 DRAWING AND PART NUMBERS...

Page 20: ...20 SL 2V Explosion...

Page 21: ...1 4 ATS 5171 Sight Port Nut 1 5 ATS 6110 Box 1 6 ATS5 410 Lock Nut 1 7 ATS8 544 End Nut O Ring seal quartz sleeve 1 8 ATS 7091 Gaynor Connector 1 9 ATS5 409 Brass End Nut 1 10 ATS 6111 Cover 1 11 ATS1...

Page 22: ...22 SL 8V Explosion...

Page 23: ...Retainer 8 gpm 1 6 ATS F08 8 gpm Flo Et 1 7 ATS 6110 Box 1 8 ATS5 410 Lock Nut 1 9 ATS8 544 End Nut O Ring seal quartz sleeve 1 10 ATS 7091 Gaynor Connector 1 11 ATS5 409 Brass End Nut 1 12 ATS 6111 C...

Reviews: