background image

     

2

SAFETY INSTRUCTIONS

WARNING - to guard against injury, basic safety precautions should be observed,  

including the following:

1.  READ AND FOLLOW ALL SAFETY INSTRUCTIONS.

2.  DANGER

 - To avoid possible electric shock, special care should be taken since water is present 

near electrical equipment.  Unless a situation is encountered that is explicitly addressed by the 

provided maintenance and troubleshooting sections, do not attempt repairs yourself, refer to an 

authorized service facility.

3.  Carefully examine the disinfection system after installation.  It should not be plugged in if there 

is water on parts not intended to be wet.

4.  Do not operate the disinfection system if it has a damaged power cord or plug, if it is malfunc-

tioning or if it is dropped or damaged in any manner.

5.  Always disconnect water flow and unplug the disinfection system before performing cleaning 

or maintenance activities.  Never yank the power cord to remove it from an outlet.  Grasp the 

plug and pull to disconnect.

6.  Do not use this disinfection system for other than the intended use (potable water applications.)  

The use of attachments not approved, recommended or sold by the manufacturer / distributor 

may cause an unsafe condition.

7.  Intended for indoor use.  Do not install this disinfection system where it will be exposed to the 

weather.  Do not store this disinfection system where it will be exposed to temperatures below 

freezing unless all the water has been drained from it and the water supply has been discon-

nected.

8.  Read and observe all the important notices and warnings on the water disinfection system.

9.  If an extension cord is necessary, a cord with a proper rating should be used.  A cord rated for 

less Amperes or Watts than the disinfection system rating may over heat.  Care should be taken 

to arrange the cord so that it will not be tripped over or accidentally pulled from the outlet.

1.  SAVE THESE INSTRUCTIONS.

WARNING:  

The light given off by this unit can cause serious burns to unprotected eyes and skin.  

Never look directly at a lit UV lamp.  When performing any work on the UV Disinfection System, 

always unplug the unit first.  Never operate the UV system while the lamp is outside of the UV 

chamber.

WARNING:  

The UV lamp inside of the disinfection system is rated at an effective life of approx-

imately 9,000 hours.  To ensure continuous water treatment, replace the UV lamp  

annually with the appropriate Aqua Treatment Services UV lamp.  Failure to comply may  

present a fire hazard. 

Summary of Contents for Safe Lite SL-2V

Page 1: ...Models SL 2V 8V It is important that those responsible for the installation of this equipment as well as the owner operator read this manual and carefully follow the instructions and guidelines Revis...

Page 2: ...r sold by the manufacturer distributor may cause an unsafe condition 7 Intended for indoor use Do not install this disinfection system where it will be exposed to the weather Do not store this disinfe...

Page 3: ...Flow rate must not exceed rated capacity of the unit DESCRIPTION OF EQUIPMENT The Safe Lite series are of unique design with an ultraviolet germicidal lamp housed within a single quartz sleeve surroun...

Page 4: ...personnel look at a U V lamp in operation EXCEPT through an external Site Port lens located on the outside of the unit 3 U V disinfection units must always be properly grounded INSTALLATION The Safe L...

Page 5: ...selected Simply hand press or slightly tap in the flow control until it sits on the inside ledge of the machined port 3 Using the mounting box provided secure unit to wall or other surface Make sure t...

Page 6: ...eposit onto the quartz sleeve After sufficient film has formed on the quartz sleeve the ability of the ultraviolet germicidal rays to pass though the quartz sleeve and into the water may be impaired T...

Page 7: ...shortwave U V to effectively kill bacteria Failure to replace the U V lamps every 9 000 hours may cause bacteriological breakthrough Should the use of the disinfection unit be intermittent in no case...

Page 8: ...o determine how much chlorine solution should be used In some instances removing the seal to measure the water can be a difficult task and it is easier to guess at the quantity of disinfectant needed...

Page 9: ...w to protect wells and springs and keep them from becoming contaminated call or visit your local Cooperative Extension office or your nearest certified water treatment specialist TABLE 1 Quantity of s...

Page 10: ...ater System Using Well Sanitizer Pellets Table 1 Well Diameter Inches Weight of Pellets lbs oz Cups of Pellets Number of Pellets To produce a 400 P P M chlorine dosage NOTE Pellets Weight 1 14 gram ea...

Page 11: ...n water back down the well this also rinses upper portion of well After about 15 minutes of recirculation the water a strong chlorine odor should be apparent Turn off hose 6 Bypass water softener and...

Page 12: ...gulator or above 120 volts Line cord disconnected Check replace or outlet defective Defective U V lamp Replace Defective lamp ballast Check output voltage Replace ballast Loose open circuit wire Trace...

Page 13: ...proval by Aqua Treatment Service Inc the Dealer Distributor should return the item to Aqua Treatment Service Inc 194 Hempt Road Mechanicsburg PA 17050 prepaid If a part proves to be defective it will...

Page 14: ...120V 03 03 AMPS 120V 25 25 INLET OUTLET SIZE 3 8 STD MNPT 3 4 STD MNPT TOTAL DIMENSION INCL WALL MOUNT L X W X D 15 5 X 5 X 5 21 X 5 X 5 DISINFECTION CHAMBER MATERIAL WETTED PARTS 304 SS TUBING 304 SS...

Page 15: ...bottle or sterile container being used Holding the cap in a down position the operator should then hold his breath while taking the sample so as to avoid oral contamination of the sample The operator...

Page 16: ...16 WIRING DIAGRAM...

Page 17: ...17...

Page 18: ...18...

Page 19: ...19 DRAWING AND PART NUMBERS...

Page 20: ...20 SL 2V Explosion...

Page 21: ...1 4 ATS 5171 Sight Port Nut 1 5 ATS 6110 Box 1 6 ATS5 410 Lock Nut 1 7 ATS8 544 End Nut O Ring seal quartz sleeve 1 8 ATS 7091 Gaynor Connector 1 9 ATS5 409 Brass End Nut 1 10 ATS 6111 Cover 1 11 ATS1...

Page 22: ...22 SL 8V Explosion...

Page 23: ...Retainer 8 gpm 1 6 ATS F08 8 gpm Flo Et 1 7 ATS 6110 Box 1 8 ATS5 410 Lock Nut 1 9 ATS8 544 End Nut O Ring seal quartz sleeve 1 10 ATS 7091 Gaynor Connector 1 11 ATS5 409 Brass End Nut 1 12 ATS 6111 C...

Reviews: